Key result
Higher 30-day mean gradient after mitral valve-in-valve implantation linked to ~31% increased 1-year major events.
Why the study?
The clinical impact of residual gradient and the value of intraprocedural 3D anatomic orifice area after transcatheter mitral valve-in-valve implantation were not well defined.
Does residual gradient and 3D orifice area impact the one-year composite of all-cause mortality, heart failure hospitalization, or re-intervention in patients undergoing transcatheter mitral valve-in-valve implantation?
Observational (n=100)
No
Does residual gradient and 3D orifice area impact the one-year composite of all-cause mortality, heart failure hospitalization, or re-intervention in patients undergoing transcatheter mitral valve-in-valve implantation?
Answers are grounded in this paper and cite their sources.
Elevated 30-day mean gradient and smaller intraprocedural 3D orifice area are associated with worse clinical outcomes and mortality after transcatheter mitral valve-in-valve implantation.
AIMS: The mitral valve-in-valve (MVIV) procedure has emerged as an important therapy in failing surgical bioprosthetic valves. We aimed to evaluate factors associated with outcome following MVIV intervention, specifically the impact of 30-day MVIV gradient. We also explored the value of intraprocedural MVIV three-dimensional anatomic orifice area (3D-AOA) by transoesophageal echocardiography in a subset of patients (n = 68). METHODS AND RESULTS: Consecutive MVIV patients from a single institution with 30-day transthoracic echocardiography (TTE) were included (N = 100). Clinical and echocardiographic variables were evaluated. The primary outcome was one-year composite of all-cause mortality, heart failure hospitalization or re-intervention. Multivariable analysis was performed to determine predictors of primary outcome. Mean age was 77.3 ± 10.6 years and pre-intervention mean mitral gradient was 11.5 ± 4.0 mmHg. Thirty-day MVIV mean gradient was 7.4 ± 2.6 mmHg with ≤1+ residual regurgitation in 99.0% of patients. Multivariable analysis identified MVIV mean gradient as the only independent determinant of the primary outcome (HR 1.31, CI: 1.07-1.61, P = 0.009). MVIV 3D-AOA was associated with a 30-day MVIV mean gradient of >7 mmHg by TTE (ROC-AUC 0.8, P < 0.001), and patients with 3D-AOA > 2 cm2 had significantly lower 1-year all-cause mortality (2.5% vs. 18.7%, Kaplan-Meier log-rank P = 0.03). CONCLUSION: Elevated 30-day mean gradient is associated with worse outcomes after MVIV, and smaller intraprocedural MVIV 3D-AOA is associated with a higher 30-day mean gradient and worse mortality. Optimizing MVIV orifice area at the time of procedure may improve valve haemodynamics and patient outcomes.
No takes yet. Share an insight, caveat, or question.
Trpkov et al. (2025) conducted an observational in Failing surgical bioprosthetic valves (n=100). 30-day MVIV mean gradient was evaluated on One-year composite of all-cause mortality, heart failure hospitalization or re-intervention (HR 1.31, 95% CI 1.07-1.61, p=0.009). Higher 30-day mean gradient after mitral valve-in-valve implantation increased the 1-year risk of mortality, heart failure hospitalization, or re-intervention.
Synapse has enriched 10 related papers. Consider them for comparative context: