Key result
Advanced cone-and-plate reactor successfully delivers uniform shear stress and up to 195 mmHg hydrostatic pressure.
Why the study?
Shear stress magnitude varies with endothelial cell location in the bloodstream and modulates their function, morphology, and gene expression, requiring improved in vitro application methods.
The described cone-and-plate reactor provides a controlled in vitro environment to study the effects of shear stress and hydrostatic pressure on endothelial cells.
Improved reactor enables uniform shear stress on endothelial cells in vitro; leaves open clinical translation pending human validation.
Endothelial cells (ECs) are permanently exposed to the blood flow and the resulting shear stress, its magnitude varying with the EC site in the blood stream. Along with other mechanical stimuli like vessel wall stretching or hydrostatic blood pressure, this shear stress modulates the endothelial cell function, morphology and gene expression. Here, we describe our improved cone-and-plate reactor that applies up to 10 dyn/cm(2) uniform wall shear stress on a defined, ring-shaped region on a culture dish. At the same time, a hydrostatic pressure of up to 195 mmHg can be applied by increasing the atmospheric pressure in the incubator box. Gas composition can be controlled additionally, used for maintaining CO2-homeostasis or inducing hypoxic conditions. For better comparability, six cone-and-plate systems can be used at the same time at different rotational velocities. The effects on cell morphology, cytoskeleton and cell alignment can be monitored during application using a laser scanning microscope. Flow conditions have been studied and a sufficient area of uniform wall shear stress could be shown. To exceed 10 dyn/cm2, we suggest an increase in medium viscosity.
No takes yet. Share an insight, caveat, or question.
Dreyer et al. (2011) studied this question. Cone-and-plate reactor was evaluated on Flow conditions and uniform wall shear stress area. An improved cone-and-plate reactor successfully applied up to 10 dyn/cm2 uniform wall shear stress and up to 195 mmHg hydrostatic pressure on endothelial cells in vitro.