Key result
FOLDAVALVE demonstrates successful implantation, repositioning, and short-term function in an ovine model.
Why the study?
Minimising delivery catheter size in TAVI while maintaining repositionability and retrievability remains a technological challenge.
The FOLDAVALVE 14 Fr transcatheter aortic valve demonstrated proof-of-concept feasibility for implantation, repositioning, and retrieval in a preclinical ovine model.
Supports repositionable TAVI development; leaves open human safety, efficacy, and durability.
AIMS: Transcatheter aortic valve implantation (TAVI) is an emerging field with technological challenges. One major challenge is to minimise delivery catheter size to reduce vascular trauma, while maintaining features such as repositionability and retrievability. This study was designed as proof of concept of FOLDAVALVE, using a short-term ovine model. METHODS AND RESULTS: FOLDAVALVE is a fully repositionable and retrievable transcatheter aortic valve with a 14 Fr delivery system whose leaflets are excluded from the stent during crimping. A short-term ovine model (n=3) was used for transfemoral implantations at the Institut Mutualiste Montsouris in Paris. There was a smooth transition of the valve over the aortic arch followed by seamless delivery into the native ovine aortic position. Implantation was followed by repositioning, resheathing and retrieval, then implantation of a fresh prosthesis in the same animal. Transoesophageal echocardiography and fluoroscopy were used to monitor the delivery and implantation processes and to demonstrate valve functionality. Animal sacrifice and direct visualisation after explant confirmed excellent final position. CONCLUSIONS: FOLDAVALVE was successfully implanted in the aortic position of an ovine model. FOLDAVALVE is a promising technology which has been shown to be feasible in this preclinical TAVI study.
No takes yet. Share an insight, caveat, or question.
Kheradvar et al. (2015) studied Aortic stenosis (experimental model) (n=3). FOLDAVALVE was evaluated on Feasibility of implantation, repositioning, and retrieval. FOLDAVALVE was successfully implanted, repositioned, and retrieved in the aortic position of an ovine model, demonstrating procedural feasibility and good short-term valve function.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: