Key result
Cardioband direct mitral annuloplasty achieves 100% successful function and significantly reduces annular dimensions in swine.
Why the study?
Preclinical safety and feasibility of a new transcatheter direct mitral annuloplasty intervention had not been reported.
Does transcatheter direct mitral annuloplasty with Cardioband demonstrate safety, feasibility, and efficacy in reducing annular size in a healthy porcine model?
Does transcatheter direct mitral annuloplasty with Cardioband demonstrate safety, feasibility, and efficacy in reducing annular size in a healthy porcine model?
Transcatheter direct mitral annuloplasty with the Cardioband device is feasible and effectively reduces annular size in an acute porcine model, supporting further development.
Answers are grounded in this paper and cite their sources.
Feasible in acute porcine model; leaves open translation to human mitral regurgitation therapy.
AIMS: The aim of the study was to report preclinical safety and feasibility of a new transcatheter direct mitral annuloplasty intervention in an acute animal model. METHODS AND RESULTS: Twenty healthy pigs underwent Cardioband (Valtech Cardio, Or Yehuda, Israel) transcatheter implantation under intracardiac echocardiographic and fluoroscopic guidance. Through a neo inferior vena cava approach, transseptal access was arranged. The device was delivered into the left atrium using a multi-steerable catheter and fixed to the mitral annulus with multiple helix anchors. Following device cinching, reduction of annular size was evaluated. In all animals the device could be successfully implanted and displayed 100% function, with the average procedure duration and fluoroscopy times being 78±23 minutes and 27±9 minutes, respectively. In total, 246 anchors (average 12.3 per device) were delivered and optimal anchor placement was achieved in 95.1%, while inadequate anchor position (4%) and injury of the coronary sinus or atrium (0.8%) occurred in the rest. Following maximal cinching, diastolic transmitral flow velocity and coaptation lengths were markedly increased (p<0.001), whereas septolateral and intercommissural distances were significantly decreased (p<0.001), when compared to pre-contraction baseline, demonstrating efficient annular reduction by the device. CONCLUSIONS: Transcatheter direct annuloplasty with a surgical-like adjustable device is feasible in the porcine animal model. The humanised porcine model has been instrumental in demonstrating feasibility and in establishing the procedural steps.
No takes yet. Share an insight, caveat, or question.
Taramasso et al. (2016) studied Healthy (preclinical animal model) (n=20). Transcatheter direct mitral annuloplasty with Cardioband vs. Pre-contraction baseline was evaluated on Successful implantation and device function. Transcatheter direct mitral annuloplasty with Cardioband was successfully implanted with 100% function in an acute porcine model, reducing annular dimensions.
Synapse has enriched 10 related papers. Consider them for comparative context: