Key result
Multiple electrical shocks cut the defibrillation threshold ~50% versus single shocks in cardiac tissue models.
Why the study?
The effectiveness and optimal feedback algorithms for elimination of spiral waves in cardiac tissue by multiple electrical shocks were not well characterized.
Effect estimate: Ratio of threshold about 0.5
Multiple electrical shocks using feedback algorithms can eliminate spiral waves in cardiac tissue models at about half the threshold of conventional single shock defibrillation.
May support lower-energy multi-shock defibrillation in models; leaves open clinical translation of feedback algorithms.
We study numerically the elimination of a spiral wave in cardiac tissue by application of multiple shocks of external current. To account for the effect of shocks we apply a recently developed theory for the interaction of the external current with cardiac tissue. We compare two possible feedback algorithms for timing of the shocks: a "local" feedback algorithm 11 (using an external electrode placed directly on the tissue) and a "global" feedback algorithm 22 (using the electrocardiogram). Our main results are: application of the external current causes a parametric resonant drift similar to that reported in previous model computations; the ratio of the threshold of elimination of the spiral wave by multiple shocks to the threshold of conventional single shock defibrillation in our model for cardiac tissue is about 0.5, while earlier, less realistic models predicted the value about 0.2; we show that an important factor for successful defibrillation is the location of the feedback electrode and the best results are achieved if the feedback electrode or the ECG lead is located at the boundary (or edge) of the cardiac tissue; the "local" and the "global" feedback algorithms show similar efficiency.
No takes yet. Share an insight, caveat, or question.
Panfilov et al. (2000) studied Spiral waves in cardiac tissue. Multiple shocks of external current vs. Conventional single shock defibrillation was evaluated on Threshold of elimination of the spiral wave (Ratio of threshold about 0.5). Multiple electrical shocks reduced the threshold for eliminating spiral waves in a cardiac tissue model to about 0.5 compared to conventional single shock defibrillation.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: