Key result
Very high-power short-duration ablation yields similar one-year freedom from atrial tachyarrhythmias vs. AI-guided ablation.
Why the study?
Very high-power short-duration ablation with close interlesion distance strategy may improve procedural efficiency of pulmonary vein isolation compared with ablation-index-guided ablation in atrial fibrillation patients.
Does very high-power short-duration ablation improve procedural efficiency and one-year outcomes compared to ablation-index-guided ablation in patients with atrial fibrillation?
Cohort (n=83)
Does very high-power short-duration ablation improve procedural efficiency and one-year outcomes compared to ablation-index-guided ablation in patients with atrial fibrillation?
Absolute Event Rate: 60% vs 55.8%
p-value: p=0.96
Very high-power short-duration ablation for pulmonary vein isolation significantly reduces procedure and radiofrequency times compared to ablation-index-guided ablation, but yields similar one-year freedom from atrial tachyarrhythmia.
vHPSD with close ILD may improve PVI efficiency in AF; leaves open durability versus AI-guided ablation.
Background Very high-power short-duration (vHPSD) ablation with close interlesion distance (ILD) strategy may improve procedural efficiency of pulmonary vein isolation (PVI) compared with ablation-index (AI)-guided ablation. Objective To compare procedural efficiency and one-year outcomes of vHPSD versus AI-guided radiofrequency (RF) PVI in patients with atrial fibrillation (AF). Methods AF patients undergoing first-time RF PVI using vHPSD (n=40, 65% paroxysmal AF, 90W/4s, ILD<4 mm) were compared with an AI-guided cohort (n=43, 74% paroxysmal AF, 30-40W, AI targets of 380-500, ILD<6 mm). In the absence of first-pass isolation (FPI), touch-up ablation (<50W) was mandatory. Following a 30-minute waiting period, bidirectional block was confirmed and acute reconnections were re-ablated. Atrial tachyarrhythmia recurrence (≥30s) was assessed over 12 months using ECG, mobile one-lead ECG monitoring, and Holter recordings. Results Patients undergoing vHPSD had smaller 3D LA volumes than those treated with AI-guided PVI (48.5±13.5mL/m 2 vs. 60.3±19.9mL/m 2 , p=0.003). Procedure and RF times were significantly shorter with vHPSD (58.0±29.7min vs. 72.4±26.0min, p=0.02; 8.3±3.0min vs. 36.9±7.6min, p<0.001, respectively). Although FPI rates were similar, acute PV reconnection tended to occurr more often with vHPSD, mainly at the carina and left PV ridge. Ablation line length correlated with LA volume and was greater in patients without FPI. One-year freedom from atrial tachyarrhythmia was similar between vHPSD and AI (60% vs. 55.8%, log-rank p=0.96). Conclusion : PVI using vHPSD with close ILD reduced procedure and RF times but resulted in more acute PV reconnections, predominantly at the carina and left PV ridge, requiring re-ablation. One-year atrial tachyarrhythmia recurrence was similar to AI-guided ablation.
No takes yet. Share an insight, caveat, or question.
Pouderoijen et al. (2026) conducted a cohort in Atrial fibrillation (n=83). Very high-power short-duration (vHPSD) ablation vs. Ablation-index (AI)-guided ablation was evaluated on One-year freedom from atrial tachyarrhythmia (p=0.96). Very high-power short-duration ablation resulted in similar one-year freedom from atrial tachyarrhythmia compared to ablation-index-guided ablation (60% vs. 55.8%, p=0.96).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: