Key result
Performance of motion sensing filters and algorithms varied by activity, with the 1-4 Hz/integration combination best for walking at a grade and 15 Hz/threshold crossing best for bicycling.
Why the study?
Conclusions regarding specific signal processing methods for motion-based rate responsive pacemakers can be affected by device-specific features, necessitating an objective comparison of commonly used filters and algorithms.
Which signal processing filter and algorithm combination best matches acceleration-indicated rate to intrinsic heart rate during various activities in normal subjects?
Cross-Sectional (n=16)
Which signal processing filter and algorithm combination best matches acceleration-indicated rate to intrinsic heart rate during various activities in normal subjects?
The optimal signal processing method for motion-based rate responsive pacing depends on the specific physical activity being performed.
Answers are grounded in this paper and cite their sources.
Activity-specific sensor optimization may refine rate-adaptive pacing; leaves open prospective patient studies before practice change.
Pacemakers that augment heart rate (HR) by sensing body motion have been the most frequently prescribed rate responsive pacemakers. Many comparisons between motion-based rate responsive pacemaker models have been published. However, conclusions regarding specific signal processing methods used for rate response (e.g., filters and algorithms) can be affected by device-specific features. To objectively compare commonly used motion sensing filters and algorithms, acceleration and ECG signals were recorded from 16 normal subjects performing exercise and daily living activities. Acceleration signals were filtered (1-4 or 15-Hz band-pass), then processed using threshold crossing (TC) or integration (IN) algorithms creating four filter/algorithm combinations. Data were converted to an acceleration indicated rate and compared to intrinsic HR using root mean square difference (RMSd) and signed RMSd. Overall, the filters and algorithms performed similarly for most activities. The only differences between filters were for walking at an increasing grade (1-4 Hz superior to 15-Hz) and for rocking in a chair (15-Hz superior to 1-4 Hz). The only differences between algorithms were for bicycling (TC superior to IN), walking at an increasing grade (IN superior to TC), and holding a drill (IN superior to TC). Performance of the four filter/algorithm combinations was also similar over most activities. The 1-4/IN (filter [Hz]/algorithm) combination performed best for walking at a grade, while the 15/TC combination was best for bicycling. However, the 15/TC combination tended to be most sensitive to higher frequency artifact, such as automobile driving, downstairs walking, and hand drilling. Chair rocking artifact was highest for 1-4/IN. The RMSd for bicycling and upstairs walking were large for all combinations, reflecting the nonphysiological nature of the sensor. The 1-4/TC combination demonstrated the least intersubject variability, was the only filter/algorithm combination insensitive to changes in footwear, and gave similar RMSd over a large range of amplitude thresholds for most activities. In conclusion, based on overall error performance, the preferred filter/algorithm combination depended upon the type of activity.
No takes yet. Share an insight, caveat, or question.
Greenhut et al. (1996) conducted a cross-sectional in Healthy volunteers (n=16). Signal processing methods (filters and algorithms) vs. Intrinsic heart rate was evaluated on Root mean square difference (RMSd) and signed RMSd between acceleration indicated rate and intrinsic HR. Performance of motion sensing filters and algorithms varied by activity, with the 1-4 Hz/integration combination best for walking at a grade and 15 Hz/threshold crossing best for bicycling.