Key result
DDD-CLS pacing improves forced vital capacity ~10% vs DDD pacing in chronotropic incompetence.
Why the study?
Chronotropic incompetence severely limits exercise tolerance due to impaired heart rate responses, and the effect of closed-loop stimulation pacing on lung function and cardiopulmonary capacity was investigated.
Does pacemaker with closed-loop stimulation (DDD-CLS) pacing improve lung function and cardiopulmonary capacity compared to DDD pacing in patients with chronotropic incompetence?
RCT (n=32)
randomized crossover
Does pacemaker with closed-loop stimulation (DDD-CLS) pacing improve lung function and cardiopulmonary capacity compared to DDD pacing in patients with chronotropic incompetence?
In patients with chronotropic incompetence, DDD-CLS pacing provides superior physiological chronotropic support, improving both lung capacity and peak oxygen uptake compared to standard DDD pacing.
Answers are grounded in this paper and cite their sources.
CLS pacing may improve lung capacity and exercise tolerance in chronotropic incompetence; leaves open confirmation in larger trials before practice change.
Chronotropic incompetence (CI) severely limits exercise tolerance due to impaired heart rate responses. This study investigated whether pacemaker with closed-loop stimulation (DDD-CLS) pacing, which provides rate acceleration in response to exertion, could enhance lung function and cardiopulmonary capacity compared pacemaker without CLS pacing in patients with CI. This randomized crossover trial included 32 patients with CI who were compared to each CLS and DDD pacing over 2 months. Spirometry was used to assess the forced vital capacity (FVC). Cycling-based cardiopulmonary exercise testing (CPET) measured peak oxygen uptake (peak VO 2 ). DDD-CLS pacing yielded significantly higher FVC (2.2 ± 0.8 L) versus DDD pacing (2.0 ± 0.7 L), p = 0.036, a 3.2% improvement. Moreover, those in the DDD-CLS mode showed a significantly higher FEV1 of 1.8 ± 0.6 L compared to the DDD mode of 1.7 ± 0.5 L (p = 0.03). Compared with DDD pacing, DDD-CLS pacing also achieved superior peak VO 2 (14.9 ± 4.0 vs 12.2 ± 3.7 mlO 2 /kg/min, p < 0.001) and peak heart rate (106.9 ± 9.7 vs 98.3 ± 19.8 bpm, p = 0.013). DDD-CLS and DDD pacing modes showed significant correlations between FVC and peak VO 2 and between peak VO 2 and peak heart rate during CPET. Compared with DDD mode, DDD-CLS pacing provided physiological chronotropic support to improve cardiopulmonary function during exertion, which enhanced lung capacity in patients with CI.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2024) conducted an RCT in chronotropic incompetence (n=32). DDD-CLS pacing vs. DDD pacing was evaluated on forced vital capacity (FVC) (p=0.036). DDD-CLS pacing improved forced vital capacity compared to DDD pacing and increased exercise capacity in patients with chronotropic incompetence.
Synapse has enriched 10 related papers. Consider them for comparative context: