Why the study?
Does closed loop stimulation (DDD-CLS) pacing improve physiological 24-h beat-to-beat heart rate and blood pressure profiles compared to conventional fixed rate DDD pacing in patients with pacemakers?
Does closed loop stimulation (DDD-CLS) pacing improve physiological 24-h beat-to-beat heart rate and blood pressure profiles compared to conventional fixed rate DDD pacing in patients with pacemakers?
DDD-CLS pacing provides more physiological heart rate fluctuations and less pronounced drops in heart rate and diastolic blood pressure at the onset of paced rhythm compared to conventional DDD pacing.
Answers are grounded in this paper and cite their sources.
Supports DDD-CLS over DDD for physiological rate response in pacemaker patients; extends RCT evidence on RR matching to 24-h profiles.
AIMS: The aims of the study were to evaluate the 24-h beat-to-beat heart rate (RR) and blood pressure changes during closed loop stimulation (DDD-CLS) pacing and conventional fixed rate DDD pacing with respect to spontaneous activity. METHODS: We simultaneously and continuously measured beat-to-beat heart rate and blood pressure for 24 h in patients implanted with Inos2+ (Biotronik GmbH, Berlin, Germany). A randomised cross-over comparison of DDD-CLS and DDD pacing was performed by short- and long-term analyses. RESULTS: Seventeen patients (10 males, aged 46-85 years) were enroled in the study: 11 completed the protocol. The percentage of atrial stimulation was 72.87% during DDD-CLS and 38.36% in DDD (P=0.003). All patients were 100% stimulated in the ventricle. On average, the percentage increase of paced RR intervals with respect to spontaneous beats was only 7.4% in DDD-CLS but 20.1% in DDD (P=0.0001). A significant correlation between spontaneous and paced RR profiles was obtained only during DDD-CLS (r(DDD-CLS)=0.77, r(DDD)=0.23, P=0.01). Short-term analysis revealed a 3.79% reduction of the escape interval in DDD-CLS and 8.19% in DDD, and the relative fall in diastolic blood pressure was 1.14% in DDD-CLS and 3.81% in DDD. CONCLUSION: DDD-CLS provided physiological heart rate fluctuations throughout the 24-h test. The blood pressure profiles of paced and spontaneous beats were comparable. The onset of paced rhythm in DDD-CLS resulted in a less pronounced decrease in heart rate and fall in diastolic pressure than in DDD.
No takes yet. Share an insight, caveat, or question.
Quaglione et al. (2005) studied this question.