High repetition rates generate artifacts in two-dimensional experiments which can impede spectral analysis. Incomplete transverse relaxation produces a modulation of the detected signal; the resultant anomalies appear at integral multiples of the resonance offset in the F1 dimension of the final spectrum. Calculations show that the z magnetization eventually falls into a cyclic steady state which reflects the order of the phase cycle. Dummy repetitions of the pulse sequence are shown to be relatively inefficient at eliminating artifacts. Ten different permutations of the echo selection COSY phase cycle are compared; the results demonstrate that RF phase cycling is a powerful method for quenching transverse interference, thus improving the quality of the data. A sequence which cancels quadrature images in F2, do offsets, and axial peaks, selects the appropriate echo, and cancels transverse interference, all within a basic cycle of only four steps, is demonstrated. In general, a well-designed 16-step phase cycle produces less interference than its 4-step counterpart. The optimum phase cycle for the echo selection COSY experiment is based upon cyclic monotonic permutation and requires a minimum of 16 steps.
No takes yet. Share an insight, caveat, or question.
Steven L. Patt (1989) studied this question.