Key result
Biventricular direct cardiac compression increases stroke volume without raising myocardial oxygen consumption.
Why the study?
Biventricular direct cardiac compression (DCC) may support failing hearts without blood/device interface complications, but its effects on ventricular performance require evaluation.
In vitro DCC effects should not guide practice; leaves open translation to in vivo biventricular support.
BACKGROUND: Biventricular direct cardiac compression (DCC) can potentially support the failing heart without the complications associated with a blood/device interface. The effect of uniform DCC on left and right ventricular performance was evaluated in 7 isolated canine heart preparations. METHODS AND RESULTS: A computer-controlled afterload system either constrained the isolated heart to contract isovolumically or simulated hemodynamic properties of physiological ejection. Biventricular DCC was provided by a chamber surrounding the heart that allowed adjustment of the compression pressure, onset time, and duration. Through a series of ventricular preloads, the effect of DCC on the end-systolic pressure-volume relationship (ESPVR) was evaluated under isovolumic and ejecting conditions. Under both conditions, DCC shifted the ESPVR of the left and right ventricles upward by an amount approximately equal to the compression pressure. The augmentation of end-systolic pressure for each initial preload tested, however, was less under ejecting conditions, because reductions in end-systolic and end-diastolic volumes occurred with ejection. Nevertheless, the net effect was to increase stroke volume. Measurement of MO2 demonstrated that at a given ventricular volume, MO2 did not change with DCC; however, peak ventricular pressure increased substantially, so that the effective pressure-volume area increased. CONCLUSIONS: Biventricular DCC can augment end-systolic pressure with no added costs of MO2. Under ejecting conditions, this augmentation of ventricular contracting ability manifests as increases in stroke volume. Thus, DCC represents a feasible alternative form of ventricular assist, and devices that support the heart in this manner should be further explored.
No takes yet. Share an insight, caveat, or question.
Artrip et al. (1999) studied this question. Biventricular direct cardiac compression (DCC) vs. Baseline (without DCC) was evaluated on End-systolic pressure-volume relationship (ESPVR). Biventricular direct cardiac compression shifted the ESPVR upward and increased stroke volume without added myocardial oxygen costs in isolated canine hearts.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: