Key result
Independent biventricular epicardial pressure achieves normal ejection fraction in a passive heart model.
Why the study?
Devices for direct cardiac compression have not been tuned to meet the individual compression requirements of the left and right ventricles, which can differ substantially.
Does applying independent epicardial pressures to the left and right ventricles improve stroke volume and ejection fraction compared to uniform pressure in a passive human heart model?
Does applying independent epicardial pressures to the left and right ventricles improve stroke volume and ejection fraction compared to uniform pressure in a passive human heart model?
Designing a direct cardiac compression pump that applies separate pressures to the left and right ventricles is critical to maintain equivalent stroke volume for both ventricles.
May guide epicardial device pressures in models; leaves open translation to active human hearts.
Direct cardiac compression (DCC) holds enormous potential as a safe and effective means to treat heart failure patients who require long-term, or even permanent, biventricular support. However, devices developed to date are not tuned to meet the individual compression requirements of the left and right ventricles, which can differ substantially. In this paper, a systematic study examining the relationship, range, and effect of independent pressures on the left and right epicardial surfaces of a passive human heart model was performed as a means to optimize cardiac output via DCC support. Hemodynamic and tissue deformation effects produced by varying epicardial compressions were examined using finite element analysis. Results indicate that 1) designing a direct cardiac compression pump that applies separate pressures to the left and right ventricles is critical to maintain equivalent stroke volume for both ventricles, and 2) left and right ventricular epicardial pressures of 340 mmHg and 44 mmHg, respectively, are required to induce normal ejection fractions in a passive heart. This pilot study provides fundamental insights and guidance towards the design of improved direct cardiac compression devices for long-term circulatory support.
No takes yet. Share an insight, caveat, or question.
Han et al. (2019) studied Heart failure (simulated). Separate left and right ventricular epicardial pressures vs. Uniform epicardial pressure was evaluated on Epicardial pressures required to induce normal ejection fractions. Applying separate left and right ventricular epicardial pressures of 340 mmHg and 44 mmHg, respectively, is required to induce normal ejection fractions in a passive heart model.