Key result
Redundant transvenous pacemaker lead causes late severe tricuspid valve stenosis and heart failure.
Why the study?
Acquired tricuspid valve stenosis as a complication of endocardial pacemaker lead placement has not been described in veterinary literature.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
Raises caution for late tricuspid stenosis from adherent transvenous leads in dogs; leaves open optimal lead strategies and surveillance protocols.
Acquired tricuspid valve stenosis (TVS) is a rare complication of endocardial lead placement in humans that has not been described in the veterinary literature. We describe a 6.5-year-old Border terrier that was presented with right-sided congestive heart failure 5.5 years after placement of a transvenous pacemaker. Severe TVS was confirmed by cardiac ultrasonography and appeared to be secondary to excessive endocardial pacemaker lead within the right atrium. Repositioning of the lead proved impossible and subsequent postmortem examination demonstrated fusion of the tricuspid valve (TV) leaflets around the endocardial lead with associated narrowing of the tricuspid valve orifice. In addition, the loop of the endocardial lead was anchored by thick fibrous tissue to the right atrial wall. This case report suggests that if sufficient endocardial lead is left inside the heart, in dogs undergoing transvenous pacemaker therapy, the redundant lead can become adherent to the tricuspid valve apparatus and cause valvular stenosis.
No takes yet. Share an insight, caveat, or question.
Connolly et al. (2007) conducted a case report in Right-sided congestive heart failure and tricuspid valve stenosis (n=1). Transvenous pacemaker lead was evaluated on Severe tricuspid valve stenosis. Excessive transvenous pacemaker lead became adherent to the tricuspid valve apparatus in a dog, causing severe valvular stenosis and heart failure 5.5 years later.
Synapse has enriched 10 related papers. Consider them for comparative context: