Key result
MRI radiofrequency energy shows no association with device parameter changes despite minor post-scan voltage drops.
Why the study?
The risks associated with MRI in individuals with implanted cardiac devices related to radiofrequency energy and device function were not fully characterized.
Cohort (n=1,464)
Answers are grounded in this paper and cite their sources.
MRI parameters and cardiac device function show inconclusive associations; leaves open safe SAR and dB/dt thresholds for practice.
Background The risks associated with MRI in individuals who have implanted cardiac devices are thought to arise from the interaction between the implanted device and static, gradient, and radiofrequency magnetic fields. Purpose To determine the relationship between the peak whole-body averaged specific absorption rate (SAR) and change in magnetic field per unit time (dB/dt), maximum specific energy dose, imaging region, and implanted cardiac device characteristics and their function in patients undergoing MRI. Materials and Methods This prospective observational cohort study was conducted from October 16, 2003, to January 22, 2015 (https://ClinicalTrials.gov, NCT01130896). Any individual with an implanted cardiac device who was referred for MRI was included. Clinical MRI protocols without SAR restriction were used. Exclusion criteria were newly implanted leads, abandoned or epicardial leads, and dependence on a pacemaker with an implantable cardioverter defibrillator without asynchronous pacing capability. For each MRI pulse sequence, the calculated whole-body values for SAR, dB/dt, and scan duration were collected. Atrial and ventricular sensing, lead impedance, and capture threshold were evaluated before and immediately after (within 10 minutes) completion of each MRI examination. Generalized estimating equations with Gaussian family, identity link, and an exchangeable working correlation matrix were used for statistical analysis. Results A total of 2028 MRI examinations were performed in 1464 study participants with 2755 device leads (mean age, 67 years ± 15 [standard deviation]; 930 men [64%]). There was no evidence of an association between radiofrequency energy deposition, dB/dt, or scan duration and changes in device parameters. Thoracic MRI was associated with decreased battery voltage immediately after MRI (β = −0.008 V, P < .001). Additionally, right ventricular (RV) lead length was associated with decreased RV sensing (β = −0.012 mV, P = .05) and reduced RV capture threshold (β = −0.002 V, P < .01) immediately after MRI. Conclusion There was no evidence of an association between MRI parameters that characterize patient exposure to radiofrequency energy and changes in device and lead parameters immediately after MRI. Nevertheless, device interrogation before and after MRI remains mandatory due to the potential for device reset and changes in lead or generator parameters.
No takes yet. Share an insight, caveat, or question.
Rahsepar et al. (2020) conducted a cohort in Patients with implanted cardiac devices undergoing MRI (n=1,464). MRI radiofrequency energy and magnetic fields vs. null (evaluated before and immediately after MRI) was evaluated on Changes in device parameters (atrial and ventricular sensing, lead impedance, and capture threshold) associated with radiofrequency energy deposition. MRI radiofrequency energy deposition was not associated with changes in device parameters, though thoracic MRI decreased battery voltage immediately after scanning.
Synapse has enriched 10 related papers. Consider them for comparative context: