Key result
A 1.0 Hz high-pass ECG filter is linked to ~9.1-fold higher odds of false-positive ST elevation.
Why the study?
High-pass filter settings used in bedside ECG monitoring systems differ from standard 12-lead ECGs and may induce ST-segment distortion, but the magnitude and clinical impact were not quantified with modern modeling.
Do higher high-pass filter settings (≥0.5 Hz) increase the odds of false-positive ST-segment elevation compared to the standard 0.05 Hz filter in patients referred for electrophysiology study?
Observational (n=150)
Do higher high-pass filter settings (≥0.5 Hz) increase the odds of false-positive ST-segment elevation compared to the standard 0.05 Hz filter in patients referred for electrophysiology study?
Answers are grounded in this paper and cite their sources.
High-pass ECG filter settings of ≥0.5 Hz, commonly used in telemetry, cause anatomically selective and clinically significant false-positive ST-segment elevation, particularly in right precordial leads.
BACKGROUND: Standard 12‑lead electrocardiograms (ECGs) use a 0.05 Hz high-pass filter, while continuous bedside monitoring systems commonly employ higher filter settings (≥0.5 Hz) to reduce baseline wander. Building upon foundational studies of filter-induced artifact, we aimed to apply modern parametric modeling to quantify the magnitude, anatomical distribution, and clinical determinants of this phenomenon. METHODS: We conducted a repeated-measures analysis of 150 patients referred for electrophysiology study. Standard 12‑lead ECGs were analyzed using four high-pass filter settings: 0.05 Hz, 0.1 Hz, 0.5 Hz, and 1.0 Hz. Linear mixed-effects models assessed the effects of filter frequency and baseline QRS morphology on ST-segment deviation. Logistic regression was used to determine the odds of clinically significant false-positive ST elevation (≥1.0 mm). RESULTS: Increasing high-pass filter frequency produced significant, frequency-dependent ST-segment elevation, with a pronounced "ceiling effect" occurring at the standard telemetry default of 0.5 Hz. The artifact was anatomically selective, predominantly affecting the right precordial leads (maximum increase at 1.0 Hz: V1 + 0.51 mm, V2 + 0.80 mm; both p < 0.001), while limb leads were minimally affected (Lead I + 0.09 mm, Lead II +0.19 mm). An R/S ratio < 1 in the right precordial leads significantly amplified artifact magnitude at 1.0 Hz. Use of a 1.0 Hz filter independently increased the odds of false-positive ST elevation ≥1.0 mm nearly ten-fold in V1 (OR: 9.14, 95% CI: 4.16-20.11; p < 0.0001) and five-fold in V2 (OR 4.96, 95% CI 2.87-8.58; p < 0.001). CONCLUSIONS: High-pass filter settings ≥0.5 Hz induce substantial, anatomically selective ST-segment elevation, highlighting the need for caution when interpreting ST segments on bedside monitoring systems.
No takes yet. Share an insight, caveat, or question.
Barmpagiannos et al. (2026) conducted an observational in referred for electrophysiology study (n=150). 1.0 Hz high-pass filter vs. 0.05 Hz high-pass filter was evaluated on clinically significant false-positive ST elevation (≥1.0 mm) in lead V1 (OR 9.14, 95% CI 4.16-20.11, p=<0.0001). Use of a 1.0 Hz high-pass filter independently increased the odds of false-positive ST elevation ≥1.0 mm nearly ten-fold in lead V1.