Key result
Modified Perceval valve implantation technique achieves ~100% procedural success with low 30-day mortality.
Why the study?
A modified implantation technique of the Perceval sutureless aortic prosthesis was developed and its outcomes needed evaluation.
Observational (n=128)
No
Answers are grounded in this paper and cite their sources.
Modified Perceval implantation achieved high success in this series; leaves open whether it improves outcomes versus standard technique in prospective trials.
BACKGROUND: In our institution, we developed a modification of the standard implantation technique of the Perceval sutureless aortic prosthesis (LivaNova, London, United Kingdom) that involves the usage of snuggers for the guiding sutures during valve deployment. The technique has been described elsewhere. In this article, we present the results of our initial case series comprising 128 consecutive patients. METHODS: From September 2016 to June 2018, 128 patients received a sutureless Perceval prosthesis (mean age 72.2 ± 8.5 years, 66 males). The data of the patients were prospectively collected in our general database and retrospectively analyzed using the SPSS software version 25 (IBM SPSS Inc., Chicago, Illinois, United States). Primary end points were primary implantation success and 30-day mortality. All patients underwent transthoracic echocardiography on discharge. RESULTS: Our cohort consisted of real-world scenario patients with infective endocarditis, bicuspid aortic valves and one patient with left ventricular assist device. However, we had 100% primary implantation success with no need for valve explantation or paravalvular leaks. All-cause 30-day mortality was 2.3% (3 patients) with no cardiac deaths. Two patients (1.6%) had a permanent neurological deficit, three patients (2.3%) had new-onset renal dialysis, and four patients (3.1%) needed a permanent pacemaker postoperatively. CONCLUSION: To our knowledge, this is the first case series to report on a modified implantation technique of the Perceval prosthesis. The Perceval valve prosthesis produces excellent outcome in most pathologies, so that current contraindications should be revised.
No takes yet. Share an insight, caveat, or question.
Zhigalov et al. (2019) conducted an observational in Aortic valve disease (n=128). Modified Perceval implantation technique (using snuggers) was evaluated on Primary implantation success. The modified implantation technique for the Perceval sutureless aortic prosthesis achieved 100% primary implantation success and a 30-day all-cause mortality rate of 2.3%.
Synapse has enriched 10 related papers. Consider them for comparative context: