Key result
AAOL-3D yields clinically acceptable RV pacing localization, averaging ~7 mm error compared to surface mapping.
Why the study?
Previous evaluation of the 3D Automated Arrhythmia Origin Localization algorithm (AAOL-3D) for predicting earliest right-ventricular activation was limited by a small sample size, requiring validation in a larger cohort and comparison with AAOL-Surface.
Does the AAOL-3D algorithm improve localization accuracy of right ventricular pacing sites compared to AAOL-Surface in patients with premature ventricular contractions?
Observational (n=20)
Does the AAOL-3D algorithm improve localization accuracy of right ventricular pacing sites compared to AAOL-Surface in patients with premature ventricular contractions?
Both AAOL-3D and AAOL-Surface algorithms achieved clinically acceptable accuracy (<10 mm) for localizing right ventricular activation sites, with comparable performance in the anatomically complex right ventricular outflow tract.
Answers are grounded in this paper and cite their sources.
AAOL-3D matches AAOL-Surface accuracy for RV site localization; leaves open need for larger prospective validation before clinical adoption.
Background The Automated Arrhythmia Origin Localization (AAOL) algorithm predicts the location of earliest ventricular activation in real time using standard ECG and patient-specific electroanatomic mapping (EAM) geometry. We previously extended AAOL to output 3D coordinates (AAOL-3D) in a multicenter cohort; however, right-ventricular (RV) evaluation was limited by a small sample size. Objective To validate AAOL-3D for localizing RV pacing sites (earliest activation locations) in a larger RV cohort and to compare its performance with AAOL-Surface. Methods RV pacing data from patients with premature ventricular contraction were analyzed. For each site, 120-ms QRS integrals (leads III, V2, V6) and 3D EAM coordinates were extracted. Within each patient, leave-one-site-out cross-validation was performed (test site; remaining sites for training). AAOL-Surface predicted a point on the EAM surface; AAOL-3D predicted full 3D coordinates. Localization error was the Euclidean distance between predicted and true coordinates. Paired tests compared methods overall and by RV subregion (RVOT vs non-RVOT). Results A total of 20 patients with 196 sites were enrolled (9.8±1.7 sites/patient). The mean age was 47.1 years and 63.6 % were female. The common comorbidities including hypertension (25.0%), diabetes (9.4%), and heart failure (6.3 %). Overall error was 7.44±2.88 mm for AAOL-3D vs 6.80±3.46 mm for AAOL-Surface (P < 0.001). RVOT (n=57): 7.83±2.92 vs 7.39±3.48 mm (P=0.133). Non-RVOT (n=139): 7.28±2.86 vs 6.55±3.43 mm (P < 0.001). Figure 1 shows statistical results and a pacing-site localization example. Conclusion Both methods achieved clinically acceptable accuracy (<10 mm). In RVOT, AAOL-3D was comparable to AAOL-Surface, supporting feasibility of direct 3D coordinate prediction in this anatomically complex, crescent-shaped region where surface projection can introduce geometric artifacts. Further evaluation and optimization across RV subregions are indicated.
No takes yet. Share an insight, caveat, or question.
Ding et al. (2026) conducted an observational in Premature ventricular contraction (n=20). AAOL-3D algorithm vs. AAOL-Surface algorithm was evaluated on Localization error (Euclidean distance between predicted and true coordinates in mm) (p=< 0.001). AAOL-3D achieved clinically acceptable RV pacing localization error compared to AAOL-Surface overall and was comparable in the RVOT.