Key result
Novel LeesCa catheter achieves 100% device success rate during LV lead implantation.
Why the study?
Complex coronary venous anatomy increases the procedural difficulty of LV lead placement in CRT, and conventional angiographic catheters may provide limited support and incomplete vessel visualization. This study evaluated the short-term efficacy and safety of a novel balloon angiographic catheter in standard CRT procedures.
Does the LeesCa™ balloon angiographic catheter achieve a device success rate exceeding the 93% target value for coronary venography in patients with heart failure undergoing CRT left ventricular lead implantation?
Does the LeesCa™ balloon angiographic catheter achieve a device success rate exceeding the 93% target value for coronary venography in patients with heart failure undergoing CRT left ventricular lead implantation?
The novel dual-lumen LeesCa™ balloon angiographic catheter achieved a 100% device success rate and 98.97% immediate procedural success rate with no device-related adverse events during CRT left ventricular lead implantation.
Answers are grounded in this paper and cite their sources.
Exceeding target success supports procedural feasibility in CRT; leaves open randomized confirmation versus conventional catheters.
Background Cardiac resynchronization therapy (CRT) is a key treatment for heart failure with reduced ejection fraction, with optimal benefits depending on accurate placement of the left ventricular (LV) leads in ideal target veins. Complex coronary venous anatomy increases procedural difficulty. Conventional angiographic catheters may provide limited support and incomplete vessel visualization. This study aimed to evaluate the short-term efficacy and safety of a balloon angiographic catheter in standard CRT procedures. Methods This prospective, multicenter, single-arm, target-value clinical trial assessed a novel balloon angiographic catheter (LeesCa™) for coronary venography in patients undergoing LV pacing lead implantation via the coronary sinus. Ninety-seven eligible patients (73 men) were enrolled across four centers between October 28, 2024, and July 10, 2025. All patients met the indications for LV lead implantation via the coronary sinus. Coronary venography was performed using the LeesCa™ balloon angiographic catheter. The primary endpoint was the device success rate. Secondary endpoints included the immediate procedural success rate, LV lead implantation time, contrast medium volume, total fluoroscopy time, radiation dose, device performance, and safety outcomes. Patients were followed until hospital discharge or for 7 days post-procedure, whichever occurred first. Results Device success was achieved in all patients (97/97, 100%; 95% confidence interval: 96.27%–100%), which was significantly higher than the predefined target value of 93% (P < 0.001). The immediate procedural success rate was 98.97% (96/97). One patient failed to undergo trans‑coronary sinus lead implantation because of small, tortuous target veins. The mean LV lead implantation time was 18.16 ± 12.14 min, the contrast volume was 29.65 ± 24.50 mL, the fluoroscopy time was 1113.00 ± 808.44 s, and the radiation dose was 368.92 ± 344.71 mGy. Three mild adverse events (puncture-site hematoma, contrast reaction, and vasovagal reflex) were documented, resulting in an overall adverse event rate of 3.09% (3/97). All device performance scores exceeded 3 points, satisfying the requirements for clinical application. Conclusions The LeesCa™ balloon angiographic catheter can safely and effectively complete coronary venous angiography during LV pacing lead implantation, providing satisfactory angiographic visualization and stable performance. Trial registration: Chinese Clinical Trial Registry (registration number: ChiCTR2400089647; date of registration: September 12, 2024).
No takes yet. Share an insight, caveat, or question.
Zhang et al. (2026) studied Heart failure with reduced ejection fraction undergoing left ventricular lead implantation for cardiac resynchronization therapy (n=97). LeesCa™ balloon angiographic catheter vs. Predefined target success rate of 93% (single-arm study) was evaluated on Device success rate (effective coronary sinus balloon occlusion, successful retrograde coronary venography visualizing at least one LV lateral branch vein, and intact catheter withdrawal) (95% CI 96.27%–100%, p=<0.001). The LeesCa™ balloon angiographic catheter achieved a 100% device success rate, exceeding the predefined 93% target value during LV lead implantation.