Key result
Percutaneous CRT achieves ~93% implantation success in patients with abnormal coronary venous anatomy.
Why the study?
Abnormalities of the coronary venous system present challenges in delivering cardiac resynchronization therapy, motivating an investigation into these anatomical variations and methods to circumvent the resulting difficulties.
Can cardiac resynchronization therapy devices be successfully and safely implanted percutaneously in heart failure patients with substantial abnormalities of coronary venous anatomy?
Observational (n=15)
Yes
Can cardiac resynchronization therapy devices be successfully and safely implanted percutaneously in heart failure patients with substantial abnormalities of coronary venous anatomy?
Demonstrates that percutaneous CRT implantation is feasible in over 90% of patients with complex coronary venous anomalies such as persistent left SVC and coronary sinus ostial atresia when utilizing specialized techniques like levophase venography, femoral pull-through, and venoplasty, avoiding surgical epicardial lead placement.
Answers are grounded in this paper and cite their sources.
May support percutaneous CRT in rare coronary venous anomalies; leaves open need for prospective validation before wider adoption.
AIMS: To investigate the abnormalities of the coronary venous system in candidates for cardiac resynchronization therapy (CRT) and describe methods for circumventing the resulting difficulties. METHODS: From four implanting institutes, data of all CRT implants between October 2008 and October 2020 were screened for abnormal cardiac venous anatomy, defined as an anatomical variation not conforming to the accepted 'normal' anatomy. Patient demographics, procedural detail, and subsequent left ventricle (LV) lead pacing indices were collected. RESULTS: From a total of 3548 CRT implants, 15 (0.42%) patients (80% male) of 72.2 ± 10.6 years in age with an LV ejection fraction of 34 ± 10.3% were identified to have had an abnormal cardiac venous anatomy over the study period. There were 13 cases of persistent left side superior vena cava (pLSVC), five of which had coronary sinus ostium atresia (CSOA) including two with an "unroofed" coronary sinus (CS); one patient had a unique anomalous origin of the CS and one patient had an isolated CSOA. In total 14 patients (60% repeat attempt) had successful percutaneous implant under general anesthesia (46.7%) via the cephalic vein (59.1%), using the femoral approach (53.3%) for levophase venography and/or pull-through, including one case of endocardial LV implant. Pacing follow-up over 37.64 ± 37.6 months demonstrated LV lead threshold between 0.62 and 2.9 volts (pulsewidth 0.4-1.5 ms) in all cases; five patients died within 2.92 ± 1.6 years of a successful implant. CONCLUSION: CRT devices can be implanted percutaneously even in the presence of substantial abnormalities of coronary venous anatomy. Alternative routes of venous access may be required.
No takes yet. Share an insight, caveat, or question.
Akhtar et al. (2022) conducted an observational in Abnormal coronary venous anatomy in candidates for cardiac resynchronization therapy (n=15). Percutaneous cardiac resynchronization therapy (CRT) implantation using alternative venous access and techniques (e.g., femoral pull-through, levophase venography, venoplasty) was evaluated on Successful percutaneous CRT implantation. Percutaneous CRT implantation succeeded in 14 of 15 patients with abnormal coronary venous anatomy, maintaining LV lead thresholds of 0.62-2.9 V over 37.64 months of follow-up.
Synapse has enriched 10 related papers. Consider them for comparative context: