Key result
Snare-assisted retrieval successfully rescues a locked-open TEER device without requiring a second transseptal puncture.
Why the study?
Complications during transcatheter edge-to-edge repair deployment related to device malfunction are uncommon but necessitate prompt decision-making and management, such as when a clip fails to close due to mechanical locking failure.
Case Report (n=1)
No
This case report describes a novel snare-assisted bailout technique to safely close and retrieve a mechanically locked-open transcatheter mitral valve repair clip without requiring a second transseptal puncture or tearing the atrial septum.
Highlights importance of prompt decision-making for device malfunction in TEER; extends limited MyClip experience but leaves open confirmatory data.
Transcatheter edge-to-edge repair (TEER) is an established intervention for patients having severe mitral regurgitation (MR) with prohibitive surgical risk. Complications during deployment related to device malfunction are uncommon but necessitate prompt decision-making and management. A 54-year-old gentleman with dilated cardiomyopathy, severe left ventricular dysfunction, and severe MR underwent TEER using the MyClip system (Meril Life Sciences Pvt. Ltd., Vapi, Gujarat, India). Initial deployment across the A2-P2 scallops of mitral leaflets achieved partial MR reduction. Significant residual MR required a second clip for optimal leaflet coaptation. While deploying the second clip, it failed to close because of a mechanical locking failure. So, we everted the clip, pulled back to LA, and it stayed in open position and failed to close. We report a novel bailout snare technique that enabled safe retrieval without the need for a second transseptal puncture. The ability to close the clip arms prior to withdrawal minimized the risk of tearing the atrial septum and facilitated the successful completion of the TEER procedure. Another clip was deployed successfully, resulting in significant MR reduction. This case demonstrates a rare complication during deployment, a malfunction of the device, and a novel strategy for safely bailing out a challenging intra-procedural complication.
No takes yet. Share an insight, caveat, or question.
Kaliyamoorthy et al. (2026) conducted a case report in Severe mitral regurgitation with dilated cardiomyopathy and severe left ventricular dysfunction (n=1). Snare-assisted bailout retrieval of a mechanically locked-open MyClip TEER device was evaluated on Safe retrieval of the locked-open TEER clip and procedural success. Snare-assisted retrieval safely closed and removed a locked-open MyClip TEER device in 1 patient (54-year-old male) without a second transseptal puncture, enabling successful MR reduction.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: