Key result
Omron HEM-7120 meets pediatric BP accuracy criteria with a ~4 mmHg mean systolic difference.
Why the study?
The accuracy of the Omron HEM-7120 automated oscillometric upper-arm home blood pressure monitor in children and adolescents against the AAMI/ESH/ISO Universal Standard numerical criteria was evaluated.
Does the Omron HEM-7120 automated oscillometric upper-arm home blood pressure monitor meet the numerical accuracy criteria of the AAMI/ESH/ISO 81060-2:2018 Universal Standard compared with mercury sphygmomanometry in children and adolescents?
Cross-Sectional (n=85)
Two observers were blinded to each other's readings and to the test-device display
No
Does the Omron HEM-7120 automated oscillometric upper-arm home blood pressure monitor meet the numerical accuracy criteria of the AAMI/ESH/ISO 81060-2:2018 Universal Standard compared with mercury sphygmomanometry in children and adolescents?
The Omron HEM-7120 met the numerical accuracy thresholds of the AAMI/ESH/ISO Universal Standard in children and adolescents, but protocol deviations prevent full formal validation and unrestricted clinical recommendation.
Should not yet change pediatric BP monitoring practice; leaves open need for protocol-compliant validation studies in children.
This study evaluated the Omron HEM-7120 automated oscillometric upper-arm home blood pressure monitor in children and adolescents against the numerical accuracy criteria of the AAMI/ESH/ISO Universal Standard. Test readings were compared with mercury auscultatory readings from two physicians using the same-arm sequential method. Three cuffs covered arm circumferences of 12-18, 17-22, and 22-32 cm. Of 90 participants, 85 were analyzed, yielding 255 comparisons. Mean differences were -4.0 ± 5.94 mmHg for systolic and -1.8 ± 7.20 mmHg for diastolic blood pressure, meeting Criterion 1. Participant standard deviations were 5.07 and 6.48 mmHg, below Criterion 2 limits. However, the requirements regarding age, reference-blood-pressure and cuff distributions, reference-measurement variability, and observer qualification were not fully met. The device met numerical criteria in this cohort, but it is not fully validated under the Universal Standard.
No takes yet. Share an insight, caveat, or question.
Pisithkul et al. (2026) conducted a cross-sectional in Blood pressure measurement validation in children and adolescents (n=85). Omron HEM-7120 automated oscillometric upper-arm home blood pressure monitor vs. Calibrated mercury sphygmomanometer (Diplomat-presameter) simultaneous dual-observer auscultation was evaluated on Mean device-minus-reference systolic blood pressure (SBP) difference across 255 paired comparisons (AAMI/ESH/ISO Universal Standard Criterion 1) (MD -4.01 mmHg (SD 5.94 mmHg) for SBP; MD -1.83 mmHg (SD 7.20 mmHg) for DBP). The Omron HEM-7120 met AAMI/ESH/ISO numerical accuracy criteria in children with mean differences of -4.01±5.94 mmHg for SBP and -1.83±7.20 mmHg for DBP, though protocol requirements were not fully met.