Key result
The Omron 705 IT oscillometric device accurately measured blood pressure in normotensive children and adolescents, with mean systolic and diastolic differences of 4.0±4.8 mmHg and -2.1±5.9 mmHg.
Why the study?
Does the Omron 705 IT oscillometric device accurately measure blood pressure compared to mercury sphygmomanometers in normotensive children and adolescents?
Population
197 normotensive children and adolescents (99 boys, mean age 10.6 ± 2.4 years, range 6-16 years)
Comparison
Blood pressure measurement using the Omron 705… vs Simultaneous blood pressure measurements by two…
Design
Cross-sectional
Authors
Loading...
Supports device accuracy in normotensive youth; leaves open validation in hypertensive and broader pediatric populations.
Cross-Sectional (n=197)
Does the Omron 705 IT oscillometric device accurately measure blood pressure compared to mercury sphygmomanometers in normotensive children and adolescents?
Effect estimate: Mean SBP diff 4.0±4.8 mmHg, DBP diff -2.1±5.9 mmHg
The Omron 705 IT oscillometric device is accurate for measuring blood pressure in normotensive children and adolescents.
Stergiou et al. (2006) conducted a cross-sectional in Normotensive children and adolescents (n=197). Omron 705 IT oscillometric device vs. Mercury sphygmomanometers was evaluated on Absolute device-observer blood pressure differences (Mean SBP diff 4.0±4.8 mmHg, DBP diff -2.1±5.9 mmHg). The Omron 705 IT oscillometric device accurately measured blood pressure in normotensive children and adolescents, with mean systolic and diastolic differences of 4.0±4.8 mmHg and -2.1±5.9 mmHg.