Key result
Superior RA leadless pacing succeeds in ~57% of attempts and better preserves P-wave duration vs RAA.
Why the study?
Does atrial leadless pacemaker implantation at the superior right atrium improve physiological atrial activation while maintaining feasibility and safety compared to right atrial appendage base pacing in patients undergoing dual-chamber leadless pacemaker implantation?
Cohort (n=68)
Does atrial leadless pacemaker implantation at the superior right atrium improve physiological atrial activation while maintaining feasibility and safety compared to right atrial appendage base pacing in patients undergoing dual-chamber leadless pacemaker implantation?
Implanting an atrial leadless pacemaker at the superior right atrium is feasible with operator experience and better preserves physiological atrial activation and AV conduction compared to right atrial appendage base pacing.
Superior RA leadless pacing may better preserve atrial activation than RAA; hypothesis-generating and requires randomized confirmation before adoption.
Right atrial appendage (RAA) pacing may produce nonphysiological atrial activation. To evaluate the feasibility, safety, and electrocardiographic characteristics of atrial leadless pacemaker implantation at the superior right atrium (RA). Sixty-eight consecutive patients undergoing dual-chamber leadless pacemaker implantation were studied. Procedural, device, electrocardiographic, and clinical outcomes were compared between superior RA and RAA base pacing. Superior RA implantation succeeded in 26/46 attempts (56.5%), increasing from 27.8% to 60.0% and 92.3% across sequential periods. In noncrossover cases, procedure (72.7±16.4 vs 76.8±27.2 minutes; P=0.533) and fluoroscopy times (15.4±5.8 vs 18.6±11.9 minutes; P=0.249) were comparable. P-wave duration remained unchanged with superior RA pacing (99.7±29.9 to 98.1±25.6 ms; P=0.543) but increased with RAA pacing (106.7±23.1 to 134.9±25.6 ms; P<0.001); P-wave terminal force in V1 changed from 3.6% to 3.6% versus 10.0% to 45.0%, respectively (P<0.001 for RAA). PR intervals increased from 228.4±72.8 to 229.7±77.5 ms (P=0.038) and from 205.5±72.0 to 234.4±72.6 ms (P<0.001), respectively; the increase was smaller with superior RA pacing (+5.7±11.7 vs +38.6±31.3 ms; P<0.001). At 3 months, sensing amplitudes (3.4±1.5 vs 2.4±1.1 mV; P=0.091) and pacing thresholds (0.8±0.5 vs 0.8±0.3 V; P=0.785) were comparable. No pericardial tamponade occurred. During 10.4±4.6 months, mortality, heart failure events, and atrial fibrillation incidence did not differ significantly. Superior RA pacing was feasible, showed a marked learning curve, and better preserved intrinsic atrial activation than RAA pacing.
No takes yet. Share an insight, caveat, or question.
Ogano et al. (2026) conducted a cohort in Indication for dual-chamber leadless pacemaker implantation (n=68). Atrial leadless pacemaker implantation at the superior right atrium (RA) vs. Right atrial appendage (RAA) base pacing was evaluated on Feasibility (implantation success), electrocardiographic characteristics (P-wave duration and PR interval changes), and safety/clinical outcomes (p=<0.001). Superior RA leadless pacing succeeded in 56.5% of attempts and preserved P-wave duration (99.7 to 98.1 ms; P=0.543) better than RAA pacing (106.7 to 134.9 ms; P<0.001).