Key result
Optimized filtering PPG extraction closely mirrors ECG-PPG phase synchronization with ~6% absolute difference.
Why the study?
This study was conducted to compare four real-time methods of extracting heartbeat intervals from photoplethysmogram signals to estimate the total percentage of phase synchronization of autonomic circulatory regulation circuits.
Does extracting cardiointervalograms from a univariate photoplethysmogram signal accurately estimate the total percentage of phase synchronization of autonomic circulation regulation compared to simultaneous ECG and PPG recordings in healthy volunteers and patients with COVID-19?
Does extracting cardiointervalograms from a univariate photoplethysmogram signal accurately estimate the total percentage of phase synchronization of autonomic circulation regulation compared to simultaneous ECG and PPG recordings in healthy volunteers and patients with COVID-19?
Demonstrates that autonomic circulatory phase synchronization can be reliably estimated using a single PPG sensor (via dual-band filtering to detect local minima) without requiring simultaneous ECG, enabling implementation in compact, wearable diagnostic devices.
Answers are grounded in this paper and cite their sources.
Lowest-error dual-band PPG filtering supports single-sensor phase synchronization assessment; leaves open prospective validation in wearable devices.
In this paper, we compare four methods of extracting the sequence of intervals between heartbeats from the photoplethysmogram signal to estimate the total percentage of phase synchronization of autonomic regulation circuits of blood circulation that are designed to operate in real time. In the analysis of the experimental ensemble of records of healthy subjects, the optimal parameters for the used approaches were selected. We compared cardiointervalograms obtained from photoplethysmogram signals with cardiointervalograms isolated from simultaneously recorded electrocardiogram signals during the analysis of phase synchronization between the phases of the extracted signals in the low-frequency range (0.04–0.15 Hz), which reflects the element activity of autonomic control of blood circulation. The operability of the approaches used in the analysis of the total percentage of phase synchronization of the autonomic regulation circuits of the cardiovascular system based on the photoplethysmogram signal for groups of healthy people and patients during COVID-19 was demonstrated.
No takes yet. Share an insight, caveat, or question.
Kurbako et al. (2022) studied Healthy status and COVID-19 infection (n=46). Univariate photoplethysmogram (PPG) signal cardiointervalogram extraction (Methods 1-4) vs. Simultaneously recorded electrocardiogram (ECG) and photoplethysmogram (PPG) signal analysis (reference method) was evaluated on Absolute difference in total percentage of phase synchronization (S) between univariate PPG methods and the reference ECG+PPG method. Method 4 for PPG cardiointervalogram extraction achieved the lowest mean absolute difference in phase synchronization compared to reference ECG-PPG.