Key result
Novel dark-blood saturation-recovery imaging measures healthy right ventricular free wall T1 at ~1130 ms.
Why the study?
Right ventricular T1 quantification is desirable for managing congenital heart disease and pulmonary hypertension, but it is technically difficult due to the thin, trabeculated, mobile wall, adjacent blood and epicardial fat, and proximity to sternal wires.
Observational (n=5)
No
Answers are grounded in this paper and cite their sources.
Shows the feasibility of free-breathing dark blood saturation-recovery T1 mapping in healthy volunteers, highlighting moderate reproducibility alongside myocardial T1 underestimation.
Right ventricular (RV) T1 quantification is desirable in managing congenital heart disease and pulmonary hypertension patients where RV fibrosis is implicated. RV T1 quantification is technically difficult because of the thin trabeculated mobile wall of complex geometry, impacted by adjacent blood and epicardial fat, plus proximity to sternal wires in some cases. Prior work has measured RV T1 by multi-shot segmented imaging , further by IDEAL fat-water separation in SASHA extended to suppress blood signal by inflow of saturated blood. We present initial results by single-shot imaging with motion-corrected (MoCo) averaging aiming to: 1) reduce ghost artifacts arising in a segmented scan, 2) apply fat-water separation, 3) null blood within the RV, 4) facilitate anchor image acquisition and 5) permit free-breathing acquisition. Data for five healthy volunteers was acquired during free-breathing (FB) on a 1.5T Siemens Avanto by MoCo averaging of fat-water separated single-shot scans. Dark blood motion-sensitized preparation (MSPrep) parameters and feasibility were investigated. Saturation recovery sampling was applied to FB acquisitions of a single short-axis slice across the RV free wall. The sampling comprised four 20-cycle scans at Ts≈600ms and two anchor scans at Ts>6sec (20 cycles acquired with long recovery gaps), all repeated three times per subject for intra-session reproducibility (Figure 1 ). Complex MoCo averaging was configured at fixed 50% acceptance i.e. the 10 shots at most similar respiratory phase. Imaging parameters were: TE=1.0, 2.7, 4.3ms, FA 20°, FOV 360x270mm, 6x1.9x2.5mm acquired voxels, TGRAPPA rate 3, requiring a diastolic shot duration 190ms. Normal volunteer RV T1 by fat-water separated, MSPrep dark blood imaging: A) MoCo averaged water-only image at Ts 600ms, B) MoCo averaged water-only anchor image at same window/level, C) MoCo averaged fat only image, D) T1 map generated from registration and 2-point fit of the six images per sampling scheme. Optimised BIR-4 saturation efficiency η was <0.5% of M 0 over typical 1.5T cardiac B 0 and B 1 distortion. The MSPrep aimed to null blood signal by through-slice velocity-sensitized dephasing without losing myocardial signal, by adjustable time of application and velocity-sensitivity Field-of-Speed . Mean RV free wall and septal T1s were independently measured by two observers from 2-parameter fit pixelwise maps assuming η=1. Moderate free-breathing reproducibility of RV T1 was demonstrated (Table 1 ). There was usually an underestimate of myocardial T1 and often a currently unexplained non-uniformity of T1 across the heart. The MSPrep generally nulled blood when applied typically 50ms after the start of diastasis, with FoS 10-25cm/s. However, subject-specific optimisation of both MSPrep parameters was necessary to null blood while avoiding myocardial signal loss. Although FB RV T1 quantification is feasible with the proposed method, further technical development work is required and underway towards improved precision and accuracy.
No takes yet. Share an insight, caveat, or question.
Heng et al. (2015) conducted an observational in Healthy volunteers (pilot imaging for right ventricular myocardial T1 quantification) (n=5). Free-breathing fat-water separated dark blood saturation-recovery imaging with motion-corrected averaging was evaluated on Free-breathing right ventricular free wall myocardial T1 (ms). Free-breathing fat-water separated dark blood saturation-recovery imaging yielded mean RV free wall T1 of 1130±62 ms and 1139±61 ms across two observers in healthy volunteers.