Key result
Monopolar electrocautery triggers ~9 inappropriate ICD shocks during hip arthroplasty with an umbilical pad.
Why the study?
Current recommendations focus on electromagnetic interference from monopolar electrocautery during supraumbilical surgeries, while infraumbilical surgeries are not considered high risk and do not require routine magnet application to prevent inappropriate ICD therapy.
Case Report (n=1)
No
Answers are grounded in this paper and cite their sources.
May trigger inappropriate ICD therapies with umbilical electrocautery pad in hip arthroplasty; leaves open optimal pad placement in ICD patients.
Electromagnetic interference (EMI) is a known risk factor for triggering inappropriate therapy from implantable cardioverter-defibrillators (ICDs). Recommendations from the American Society of Anesthesiologists focus on EMI when using monopolar electrocautery for supraumbilical surgeries. Infraumbilical surgeries are not considered high risk for EMI; thus, no magnet must be applied routinely to prevent inappropriate ICD therapy intraoperatively. We describe a case of a 71-year-old woman who presented for left total hip arthroplasty with a history of an ICD. The patient's history was significant for non-ischemic cardiomyopathy. Monopolar electrocautery was used, and the level of the surgery was below the umbilicus. She experienced nine inappropriate ICD therapies intraoperatively with no long-term sequelae. The location of the electrocautery dispersion pad may have contributed to inappropriate therapies. Thus, the dispersion pad location should be considered when deciding whether to suspend anti-tachycardia functions intraoperatively. We present a case of inappropriate therapy from an ICD and make a recommendation for preventing such events.
No takes yet. Share an insight, caveat, or question.
DeLeon et al. (2023) conducted a case report in Non-ischemic cardiomyopathy with an implantable cardioverter-defibrillator undergoing left total hip arthroplasty (n=1). Monopolar electrocautery with return electrode at the umbilicus level without magnet application was evaluated on Inappropriate intraoperative implantable cardioverter-defibrillator (ICD) therapies. Monopolar electrocautery during total hip arthroplasty with an umbilical-level dispersion pad triggered 9 inappropriate ICD shocks (30 J each) and 4 anti-tachycardia pacing events.
Synapse has enriched 10 related papers. Consider them for comparative context: