Key result
Deeper TAVR implantation reduces diastolic left valve-to-coronary distance ~27% in biphasic digital twin modeling.
Why the study?
Coronary occlusion is a lethal complication of TAVR that still occurs in 14.3% of patients meeting standard static computed tomography thresholds (ostial height >12 mm), prompting the development of a biphasic digital twin to assess dynamic coronary occlusion risk.
Does a biphasic digital twin integrating systolic and diastolic ECG-gated CT angiography improve the assessment of dynamic coronary occlusion risk compared to static CT assessment in a patient with severe aortic stenosis undergoing TAVR?
Case Report (n=1)
Does a biphasic digital twin integrating systolic and diastolic ECG-gated CT angiography improve the assessment of dynamic coronary occlusion risk compared to static CT assessment in a patient with severe aortic stenosis undergoing TAVR?
A biphasic digital twin integrating systolic and diastolic CT angiography captured dynamic aortic root motion, leaflet expansion, and implantation-depth-dependent changes in valve-to-coronary distance in a TAVR patient who experienced delayed coronary occlusion not predicted by static coronary height.
Answers are grounded in this paper and cite their sources.
Dynamic modeling shows depth-dependent VTC trade-offs missed by static CT; hypothesis-generating for biphasic digital twins in TAVR planning.
Coronary occlusion (CO) is a lethal complication of transcatheter aortic valve replacement (TAVR) (incidence, 0.35%–4.0%; in-hospital mortality, approximately 50%). Although static computed tomography thresholds (ostial height >12 mm) remain the assessment standard, CO still occurs in 14.3% of patients meeting them. We developed a biphasic digital twin (BDT) integrating systolic and diastolic electrocardiogram-gated computed tomography angiography to assess dynamic CO risk in a 71-year-old woman with severe aortic stenosis. Biphasic aortic root dynamics and bilateral virtual valve-to-coronary distance (VTC) were evaluated at two implantation depths. Ostial height varied by approximately 5%, whereas leaflet volume increased by 22.0% in diastole. Right coronary ostial displacement (4.28 mm) was 1.74-fold that of the left (2.46 mm), and the sinotubular junction rotated 8.7°. A commissural calcified mass remained stable. Implantation 5 mm deeper reduced diastolic left VTC from 8.96 to 6.54 mm (–27%) but increased systolic right VTC from 4.57 to 5.93 mm (+29%). Delayed CO occurred 30 min after extubation despite patent intra-procedural angiography; surgery confirmed calcified anchoring, stent deflection, and coronary compression. Static coronary height did not capture dynamic CO risk in this patient. BDT may support individualised, biphasic, bilateral assessment by integrating dynamic leaflet expansion, asymmetric root motion, and calcified anchoring.
No takes yet. Share an insight, caveat, or question.
Luo et al. (2026) conducted a case report in Severe aortic stenosis undergoing transcatheter aortic valve replacement (TAVR) (n=1). Biphasic digital twin (BDT) integrating systolic and diastolic ECG-gated computed tomography angiography vs. Static computed tomography anatomical assessment was evaluated on Biphasic aortic root dynamics and bilateral virtual valve-to-coronary distance (VTC). Biphasic digital twin modeling revealed 22.0% diastolic leaflet expansion and showed 5 mm deeper TAVR implantation reduced diastolic left VTC by 27%.