Key result
Dual-energy RFA and PFA successfully terminates refractory VT adjacent to an Amplatzer VSD closure device.
Why the study?
The safety of pulsed field ablation and radiofrequency ablation using the Sphere-9 catheter for ventricular tachycardia near ventricular septal defect closure devices is unknown.
Case Report (n=1)
Colocalized pulsed field and radiofrequency ablation using a dual-energy lattice-tip catheter successfully and safely treated refractory septal ventricular tachycardia adjacent to an intracardiac VSD closure device in a single patient.
Answers are grounded in this paper and cite their sources.
Dual-energy ablation may be feasible near VSD devices for septal VT; leaves open safety and efficacy in larger prospective cohorts.
BACKGROUND: Safety regarding pulsed field ablation (PFA) and radiofrequency ablation (RFA) using the Sphere-9 Catheter for ventricular tachycardia (VT) near ventricular septal defect (VSD) closure devices is unknown. CASE SUMMARY: A 64-year-old man with extensive cardiac history including VSD closure with an Amplatzer device and complete heart block status post biventricular cardioverter-defibrillator with abandoned RV lead because of fracture presented with recurrent, symptomatic VT. Electroanatomical mapping showed VT in the basal septum adjacent to the Amplatzer device. Using the Sphere-9 catheter, RFA terminated the tachycardia, and PFA was successful near the VSD closure device without apparent complication. VT was noninducible postablation. DISCUSSION: We describe procedural reasoning and considerations to optimize safe delivery of PFA and RFA using the Sphere-9 catheter for septal VT near intracardiac hardware. TAKE-HOME MESSAGE: Colocalized PFA/RFA using the Sphere-9 catheter may be an effective treatment option for persistent VT near intracardiac devices for highly selected patients.
No takes yet. Share an insight, caveat, or question.
Martinez et al. (2026) conducted a case report in Refractory ventricular tachycardia adjacent to VSD closure device (n=1). Pulsed field ablation (PFA) and radiofrequency ablation (RFA) using the Sphere-9 dual-energy lattice-tip catheter was evaluated on Termination and noninducibility of ventricular tachycardia without apparent complication. Dual-energy RFA and PFA using the Sphere-9 catheter successfully terminated refractory VT near an Amplatzer VSD closure device without complications in a 64-year-old man (n=1).
Synapse has enriched 10 related papers. Consider them for comparative context: