Key result
Cut-balloon embolization and snaring achieve complete hemostasis and fragment retrieval without complications.
Why the study?
Retained or fractured intracoronary devices are uncommon but potentially serious complications of complex percutaneous coronary intervention, and retrieval may cause vascular injury or systemic embolization requiring several bailout techniques.
Case Report (n=1)
No
Answers are grounded in this paper and cite their sources.
This case demonstrates that a cascade of complex PCI complications—including a retained coronary snare fragment, systemic device migration to the external iliac artery, and side-branch perforation refractory to balloon inflation and microcatheter coil delivery—can be resolved percutaneously through prioritized cut-balloon embolization followed by peripheral snare retrieval.
Background Retained or fractured intracoronary devices are uncommon but potentially serious complications of complex percutaneous coronary intervention. Retrieval may itself cause vascular injury or systemic embolization and can require several bailout techniques. Case presentation: A 50-year-old man was referred after a chronic total occlusion intervention at another institution had been complicated by guidewire fracture and subsequent fracture of the snare used for retrieval. The retained snare fragment extended from the right coronary artery toward the aorta. Repeat snaring and a three-wire knot technique were unsuccessful. Balloon-assisted retrieval mobilized the fragment but caused disengagement of the system; the fragment escaped into the systemic circulation and was localized to the right external iliac artery. Control coronary angiography demonstrated perforation of the side branch used for balloon anchoring. Extravasation persisted despite prolonged balloon inflation, and microcatheter delivery for coil embolization was unsuccessful. The perforated branch was therefore occluded with a cut balloon. After angiographic hemostasis, the embolized snare fragment was captured with a peripheral snare and removed through the sheath. The patient remained hemodynamically stable, developed no tamponade, and had no procedure-related complication at 3 months. Conclusion This case illustrates a stepwise percutaneous strategy for a cascade of device-related complications. Maintaining control of the foreign body, rapidly reassessing the coronary circulation after device migration, prioritizing active coronary perforation, and retaining alternative embolization options when microcatheter delivery fails were central to successful management.
No takes yet. Share an insight, caveat, or question.
Kılınç et al. (2026) conducted a case report in Retained coronary snare fragment in the right coronary artery with retrieval-related side-branch perforation and migration to the right external iliac artery (n=1). Stepwise percutaneous retrieval (balloon-assisted mobilization and peripheral vascular snare retrieval) and cut-balloon embolization of side-branch perforation was evaluated on Angiographic hemostasis of the perforated coronary side branch, percutaneous retrieval of the embolized snare fragment, and freedom from procedure-related complications at 3 months. Cut-balloon embolization and peripheral snaring achieved complete hemostasis and fragment retrieval with 0 procedure-related complications at 3 months.
Synapse has enriched 10 related papers. Consider them for comparative context: