Key result
PFO occluder retrieval and dedicated ASD occluder placement resolves acute SVC obstruction in multifenestrated ASD.
Why the study?
Transcatheter closure of multifenestrated aneurysmal ASDs is technically challenging because device selection and positioning must close all fenestrations while accommodating complex three-dimensional atrial septal anatomy. Superior vena cava obstruction during the procedure is exceedingly rare and can be easily overlooked without systematic transesophageal echocardiography evaluation.
Case Report (n=1)
No
This case demonstrates that using a PFO occluder for multifenestrated aneurysmal ASD closure can cause near-complete SVC obstruction due to its large right atrial disc, underscoring the necessity of systematic pre-release TEE evaluation of the cavoatrial junction and careful consideration of implantation geometry.
Answers are grounded in this paper and cite their sources.
Supports routine pre-release TEE to avert SVC obstruction; hypothesis-generating for device selection in complex ASD closure.
Transcatheter closure procedure of multifenestrated aneurysmal atrial septal defects (ASDs) remains technically challenging: Optimal device selection and positioning should achieve complete closure of all fenestrations while accommodating the complex three-dimensional atrial septal anatomy and its relationship to adjacent cardiac structures. Because superior vena cava (SVC) obstruction is exceedingly rare, it may easily be overlooked unless caval venous inflow is systematically evaluated by transesophageal echocardiography (TEE) after device deployment following a pre-release TEE checklist. We report the case of a 5-year-old girl with a multifenestrated aneurysmal secundum ASD. A Uni-occluder was not available; therefore, a patent foramen ovale (PFO) occluder was used for defect closure, which resulted in near-complete obstruction of the SVC by the right atrial disc of the device. The complication was recognized after positioning by intraprocedural TEE, prompting device retrieval prior to release. Subsequent closure with a dedicated ASD occluder achieved complete closure of all fenestrations without compromising caval venous return. This case highlights the importance of a concept we refer to as implantation geometry: the interplay between device-specific design, the implantation site, and complex three-dimensional septal anatomy. Recognition and careful assessment of this relationship will help optimize procedural planning and improve procedural safety in selected patients with complex, multifenestrated ASDs. Comprehensive three-dimensional anatomical assessment, appropriate device selection, and systematic TEE evaluation of the cavoatrial junction before device release are essential for safe transcatheter closure of multifenestrated ASDs-even when defect sizing appears appropriate. In particular, the geometric characteristics of PFO occluders, including their relatively large right atrial disc, should be considered when assessing the risk of SVC obstruction.
No takes yet. Share an insight, caveat, or question.
Maximilian et al. (2026) conducted a case report in Multifenestrated aneurysmal secundum atrial septal defect (n=1). Transcatheter ASD closure (initially with a PFO occluder, subsequently replaced with a dedicated ASD occluder) was evaluated on Complete closure of all ASD fenestrations without superior vena cava obstruction. Patent foramen ovale occluder deployment in a 5-year-old girl with a multifenestrated ASD caused near-complete SVC obstruction, resolved by retrieval and placement of a dedicated ASD occluder.
Synapse has enriched 10 related papers. Consider them for comparative context: