Objective The goal of this study is to develop and validate an accurate High Performance Liquid Chromatography method to determine Clopidogrel Impurity C in the related substance analysis of Rosuvastatin and Clopidogrel tablet formulations. Methods We used an ULTRON ES-OVM Chiral (150 mm × 4.6 mm, 5 µm) column to separate the substances. The mobile phase was a mix of KH2P04 buffer and ACN in a ratio of 77:23. We set the flow rate at 1.0 mL/min. Used UV detection at 220 nm. We validated the method according to ICH guidelines for parameters like linearity, specificity, LOD, LOQ, precision, accuracy, robustness, solution stability and filter compatibility. Results The method showed excellent specificity without interference from blank, placebo, or related impurities. Theoretical plates numbered 3818, and the tailing factor was 0.856. The correlation coefficient was 1.0000 for Clopidogrel Impurity C and 0.99998 for Clopidogrel. Recovery values varied from 98.5% to 100.9%. The method proved precise, robust, and suitable for routine quality control testing. Conclusion The High Performance Liquid Chromatography method we established for determining Clopidogrel Impurity C in Rosuvastatin and Clopidogrel Tablets was demonstrated to be highly sensitive, accurate, precise and reliable.