Online-SPE LC-MS/MS accurately quantifies clopidogrel with ~10 pg/mL sensitivity and prevents metabolite back-conversion.
A newly developed and validated online-SPE LC-MS/MS method provides high sensitivity (LLOQ 10 pg/mL) and prevents back-conversion for quantifying clopidogrel in human plasma.
Answers are grounded in this paper and cite their sources.
Facilitates accurate low-level clopidogrel quantification in plasma; extends validated bioanalytical methods for pharmacokinetic research.
A quantitative method for clopidogrel using online-SPE tandem LC-MS/MS was developed and fully validated according to the well-established FDA guidelines. The method achieves adequate sensitivity for pharmacokinetic studies, with lower limit of quantifications (LLOQs) as low as 10 pg/mL. Chromatographic separations were performed on reversed phase columns Kromasil Eternity-2.5-C18-UHPLC for both methods. Positive electrospray ionization in multiple reaction monitoring (MRM) mode was employed for signal detection and a deuterated analogue (clopidogrel-d 4) was used as internal standard (IS). Adjustments in sample preparation, including introduction of an online-SPE system proved to be the most effective method to solve the analyte back-conversion in clinical samples. Pooled clinical samples (two levels) were prepared and successfully used as real-sample quality control (QC) in the validation of back-conversion testing under different conditions. The result showed that the real samples were stable in room temperature for 24 h. Linearity, precision, extraction recovery, matrix effect on spiked QC samples and stability tests on both spiked QCs and real sample QCs stored in different conditions met the acceptance criteria. This online-SPE method was successfully applied to a bioequivalence study of 75 mg single dose clopidogrel tablets in 48 healthy male subjects.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2015) studied Healthy volunteers (bioequivalence study) (n=48). Clopidogrel quantified via online-SPE tandem LC-MS/MS vs. Reference clopidogrel 75 mg tablet (and methanol protein precipitation method) was evaluated on Analytical method validation (LLOQ, extraction recovery, precision, accuracy, and 24-hour ambient back-conversion stability). Online-SPE LC-MS/MS quantified plasma clopidogrel with a 10 pg/mL LLOQ, >99.5% extraction recovery, and prevented metabolite back-conversion.
Synapse has enriched 10 related papers. Consider them for comparative context: