Key result
Oscillometric MAP correlates with invasive pressure in sepsis but lacks interchangeability outside expiratory occlusion.
Why the study?
Blood pressure monitoring is central to the hemodynamic management of septic patients, yet invasive monitoring remains scarcely available in resource-limited settings, motivating a comparison of invasive and oscillometric BP under static and dynamic conditions.
Does non-invasive oscillometric blood pressure monitoring agree and concord with invasive blood pressure monitoring under static and dynamic conditions in ICU patients with sepsis or septic shock?
Observational (n=16)
No
Does non-invasive oscillometric blood pressure monitoring agree and concord with invasive blood pressure monitoring under static and dynamic conditions in ICU patients with sepsis or septic shock?
In resource-limited ICU settings without invasive monitoring, oscillometric MAP correlates well with invasive measurements in septic patients and can support clinical decision-making, though caution is warranted during norepinephrine titration and at low BP values.
Answers are grounded in this paper and cite their sources.
Requires caution with oscillometric MAP during norepinephrine titration in sepsis; leaves open interchangeability under most conditions.
Introduction: Blood pressure (BP) monitoring is central to the hemodynamic management of septic patients. Invasive monitoring remains scarcely available in resource-limited settings. Our objective was to compare invasive and oscillometric BP in septic patients, under static and dynamic conditions, at Douala General Hospital. Patients and Methods: We conducted a prospective observational study in patients aged ≥18 years admitted to the intensive care unit for sepsis or septic shock at DGH. Paired invasive and oscillometric BP measurements (systolic, SBP; diastolic, DBP; mean, MAP) were collected under static conditions and during four dynamic challenges (passive leg raising, norepinephrine titration, fluid challenge, end-expiratory occlusion test). Agreement was assessed using Pearson or Spearman correlation, and concordance using Bland-Altman analysis against AAMI/ESH/ISO criteria. A sensitivity analysis accounted for the clustering of measurements within patients. Results: We measured 189 BP pairs in 16 patients (56.3% female; mean age 53.3 ± 17.1 years). Oscillometry underestimated SBP and MAP at low values and overestimated DBP. Oscillometric BP correlated with invasive values except during norepinephrine titration; MAP showed the best correlation (r = 0.92; p < 0.001). Only MAP during the end-expiratory occlusion test met AAMI/ESH/ISO standards, a finding confirmed on repeated-measures analysis although no longer significant at the individual level after correction (r=0.47; p=0.14; k=5). Conclusion: Oscillometric BP correlated with invasive measurements in most conditions, especially for MAP, but the two methods were interchangeable only for MAP during the end-expiratory occlusion test. In the absence of invasive monitoring, oscillometric MAP can support clinical decision-making, provided its limitations are recognized, particularly during vasopressor titration.
No takes yet. Share an insight, caveat, or question.
A et al. (2026) conducted an observational in Sepsis or septic shock (n=16). Non-invasive oscillometric blood pressure monitoring vs. Invasive blood pressure monitoring was evaluated on Agreement and concordance between paired invasive and oscillometric blood pressure measurements (SBP, DBP, MAP) under static and dynamic conditions (r = 0.92, p=<0.001). Oscillometric MAP correlated with invasive measurements in septic patients, but met interchangeability standards only during the end-expiratory occlusion test.