Background/Objectives: Manufacturing and post-processing may alter the dimensions of printed surgical guides. This exploratory study examined dimensional fidelity across four design/workflow combinations derived from one maxillary implant-planning case. Methods: Four SLA-printed guides represented a 2 × 2 arrangement combining distal reinforcement, present or absent, with support removal before or after polymerization, with one specimen per combination. Fifteen inter-sleeve distances, one reference-point distance, and six sleeve diameters were evaluated separately across printing, polymerization, and steam sterilization. Physical dimensions were compared with the corresponding digital design dimensions. Analyses were descriptive. Results: Dimensional discrepancies varied between specimens and manufacturing stages. Guide 2, incorporating distal reinforcement with support removal before polymerization, had the lowest observed mean absolute inter-sleeve error after sterilization. Changes between polymerization and post-sterilization did not consistently increase absolute discrepancy from the design. Conclusions: These specimen-specific observations support further investigation of guide design and fabrication workflow. With one guide per condition, manufacturing effects cannot be separated from specimen variability, and implant-placement accuracy cannot be inferred.