Key result
Epicardial fat thickness ≥2.8 mm on CT is linked to epicardial voltage attenuation.
Why the study?
Epicardial fat mimics scar tissue during epicardial voltage mapping by lowering voltage, requiring definition of fat thickness thresholds for voltage attenuation and normal epicardial voltage criteria without fat.
Does 3-dimensional epicardial fat distribution on cardiac CT correlate with voltage attenuation on electroanatomical mapping and affect epicardial catheter ablation efficacy in patients undergoing percutaneous epicardial mapping?
Observational (n=14)
Does 3-dimensional epicardial fat distribution on cardiac CT correlate with voltage attenuation on electroanatomical mapping and affect epicardial catheter ablation efficacy in patients undergoing percutaneous epicardial mapping?
Pre-procedural cardiac CT effectively identifies epicardial fat (>=2.8 mm) that mimics scar tissue by attenuating epicardial voltage (<1.5 mV) and predicts sites (>10 mm fat) where epicardial catheter ablation will be ineffective.
Answers are grounded in this paper and cite their sources.
Epicardial fat may affect mapping accuracy and ablation efficacy; leaves open its integration into procedural planning pending higher-level evidence.
Objectives The purpose of this study was to correlate 3-dimensional distribution of epicardial fat on computed tomography (CT) with electroanatomical (EA) voltage maps obtained during percutaneous epicardial mapping in order to determine the fat thickness cut-off that results in voltage attenuation and to establish normal ventricular epicardial voltage criteria in the absence of fat. Background Epicardial fat can mimic scar tissue when epicardial voltage mapping is performed, as both result in low epicardial voltage. Cardiac CT can differentiate epicardial fat from scar or muscle on the basis of their distinct attenuations. Methods Transcutaneous epicardial mapping was performed in a consecutive series of 14 patients. A cardiac CT was performed before the procedure and a 3-dimensional image of the epicardial fat was generated and registered with the epicardial EA voltage map. Results In patients without cardiomyopathy (n = 8), a voltage ≥1.5 mV best correlated with the absence of epicardial fat. A fat thickness ≥2.8 mm resulted in voltage attenuation and best separated low voltage (<1.5 mV) from normal voltage (≥1.5 mV; sensitivity 81%, specificity 81%, area under the curve 0.85). In patients without cardiomyopathy, the low-voltage area matched well with the area of epicardial fat. In the 6 patients with nonischemic cardiomyopathy, the low-voltage area by far exceeded the area accounted for by epicardial fat; this corresponded with the presence of scar tissue. Epicardial ablations at sites with >10 mm of fat were ineffective. Conclusions Cardiac CT identifies epicardial fat that can mimic scar tissue during epicardial EA voltage mapping, which is important during epicardial mapping and ablation.
No takes yet. Share an insight, caveat, or question.
Desjardins et al. (2010) conducted an observational in Patients undergoing percutaneous epicardial electroanatomical voltage mapping and catheter ablation (8 without cardiomyopathy, 6 with nonischemic cardiomyopathy) (n=14). Epicardial fat thickness (≥2.8 mm) identified on 3-dimensional cardiac CT registered with epicardial electroanatomical voltage maps vs. Absence of epicardial fat (<2.8 mm) / nonischemic cardiomyopathy scar tissue was evaluated on Epicardial electroanatomical voltage attenuation (<1.5 mV vs normal voltage ≥1.5 mV) (AUC 0.85 (sensitivity 81%, specificity 81%)). Epicardial fat thickness ≥2.8 mm on cardiac CT caused voltage attenuation, and fat >10 mm rendered epicardial ablation ineffective.