Answers are grounded in this paper and cite their sources.
The Dräger Alcotest® 7000 is a fuel cell–based breath alcohol screening device intended for roadside use in Canada. To be designated as an Approved Screening Device under the Criminal Code of Canada, it must meet the performance requirements established by the Alcohol Test Committee (ATC) of the Canadian Society of Forensic Science. Although the Dräger Alcotest® 7000 was approved in January 2023, its performance has not been reported in the peer-reviewed literature. Two Dräger Alcotest® 7000 devices were evaluated in accordance with the ATC Equipment Standards and Evaluation Procedures effective December 18, 2018. The evaluation included assessment of result display, breath sample acceptance, calibration stability, accuracy, precision, resistance to radio frequency interference, low-temperature performance, alcohol-free subject differentiation, classification at specified blood alcohol concentration thresholds, analysis of drinking subjects, and the use of aqueous and gaseous alcohol standards. Both devices met all applicable ATC acceptance criteria. Accuracy and precision requirements were satisfied across the tested concentration range, including at specified decision thresholds. Human subject testing demonstrated good agreement with an approved evidential breath alcohol instrument, with the majority of results within ±10 mg/100 mL. These findings confirm that the Dräger Alcotest® 7000 met the ATC performance requirements for Approved Screening Devices.
No takes yet. Share an insight, caveat, or question.
Frenette et al. (2026) studied this question.