: Increasingly, computational fluid dynamics (CFD) is helping explore the impact of variables like: cannula design/size/position/flow rate and patient physiology on venovenous (VV) extracorporeal membrane oxygenation (ECMO). Here we use a CFD model to determine what role cardiac output (CO) plays and to analyse return cannula dynamics. : Using a patient-averaged model of the right atrium and venae cava, we virtually inserted a 19Fr return cannula and a 25Fr drainage cannula. Running large eddy simulations, we assessed cardiac output at: 3.5-6.5 L/min and ECMO flow rate at: 2-6 L/min. We analysed recirculation fraction (Rf), time-averaged wall shear stress (TAWSS), pressure, velocity, and turbulent kinetic energy (TKE) and extracorporeal flow fraction (EFF = ECMO flow rate/CO). : Increased ECMO flow rate and decreased CO (high EFF) led to increased Rf (R=0.98, log fit). Negative pressures developed in the venae cavae at low CO and high ECMO flow (high CR). Mean return cannula TAWSS was >10 Pa for all ECMO flow rates, with majority of the flow exiting the tip (94.0-95.8%). : Our results underpin the strong impact of CO on VV ECMO. A simple metric like EFF, once supported by clinical data, might help predict Rf for a patient at a given ECMO flow rate. The return cannula imparts high shear stresses on the blood, largely a result of the internal diameter.
No takes yet. Share an insight, caveat, or question.
Parker et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: