Key result
Garland 1-11-1 mm configuration with Field Scaling improves mapping accuracy in atrial fibrillation PFA.
Why the study?
Integration of the FARAWAVE catheter with the EnSite X 3D-EAM system is in development, and optimal display style, interelectrode spacing, and field scaling settings have not been established.
Does the Garland 1-11-1 mm configuration with Field Scaling improve geometric mapping accuracy compared to other configurations in patients undergoing pulsed field ablation for atrial fibrillation?
Does the Garland 1-11-1 mm configuration with Field Scaling improve geometric mapping accuracy compared to other configurations in patients undergoing pulsed field ablation for atrial fibrillation?
The Garland 1-11-1 mm configuration with Field Scaling optimizes geometric concordance with CT anatomy when integrating a pentaspline pulsed field ablation catheter with a 3D mapping system.
Answers are grounded in this paper and cite their sources.
May optimize PFA mapping accuracy in observational data; leaves open whether gains improve outcomes or warrant practice change.
BACKGROUND: Pulsed field ablation (PFA) is emerging as a nonthermal and myocardium-selective therapy for atrial fibrillation. Integration of the FARAWAVE™ catheter with the EnSite™ X three-dimensional electroanatomical mapping (3D-EAM) system remains in development, and the optimal display style, interelectrode spacing, and field scaling (FS) settings have not been established. OBJECTIVE: To determine the optimal visualization configuration of the FARAWAVE™ catheter for achieving high geometric accuracy in 3D mapping. METHODS: We retrospectively analyzed 25 patients with atrial fibrillation who underwent PFA using the FARAWAVE™ pentaspline catheter visualized on the EnSite™ X system. Four configurations were tested: Medusa style (2.5 mm) and Garland styles (1-17-1 mm, 1-12-1 mm, and 1-11-1 mm). CT-derived left atrial anatomies were compared with electroanatomical reconstructions using ten concordance metrics (six distances and four angles). Accuracy was assessed with mean absolute error (MAE), Scale Score, Angle Score, and Shape Similarity. RESULTS: Distance-based MAE differed significantly among configurations (p < 0.05). The Garland 1-11-1 mm + FS setting achieved the lowest MAE (4.3-12.4 mm), while Garland 1-17-1 mm had the highest (8.5-28.4 mm). Scale Score also varied significantly (p < 0.001), whereas Angle Score and Shape Similarity were comparable across configurations. CONCLUSION: The Garland 1-11-1 mm configuration with Field Scaling provided the most accurate geometric concordance with CT anatomy. Optimized catheter display and FS correction can improve mapping fidelity and procedural safety in pulsed field ablation.
No takes yet. Share an insight, caveat, or question.
Yoshihara et al. (2025) studied this question. The Garland 1-11-1 mm configuration with Field Scaling achieved the lowest mean absolute error of 4.3-12.4 mm, improving mapping accuracy in atrial fibrillation PFA.
Synapse has enriched 10 related papers. Consider them for comparative context: