Key result
Open-source 3D Slicer plugin matches 3mensio aortic annulus measurements with a clinically irrelevant ~1.2 mm difference.
Why the study?
Existing dedicated CT software for TAVI planning is limited by significant upfront and update costs, making open-source software desirable to reduce costs and increase availability.
Does an open-source 3D slicer plugin provide comparable aortic annulus measurements to the gold standard 3mensio software in TAVI cases?
Observational (n=50)
Does an open-source 3D slicer plugin provide comparable aortic annulus measurements to the gold standard 3mensio software in TAVI cases?
An open-source 3D slicer plugin provides highly comparable aortic annulus measurements to the gold standard 3mensio software, offering a potentially cost-effective alternative for TAVI sizing.
Answers are grounded in this paper and cite their sources.
May support cost-effective TAVI planning alternatives; leaves open larger prospective validation before adoption.
Introduction TAVI planning with dedicated CT software is a key component of any TAVI program. The Pie Medical 3mensio software suite is the most used system. Some competitors are also available. However, all this software is limited by very significant costs, both upfront and regarding updates. As a result, open source or more affordable software would be ideal to reduce the costs of TAVI and increasing availability. We aimed to develop an open-source software plugin for TAVI sizing, comparing its results with the gold standard, 3mensio. Methods A python plugin for the open-source software 3D Slicer was written by an Interventional Cardiologist with python coding knowledge, aided by artificial intelligence coding assistants. The software requires the input of the 3 cusps nadir and marking reference points of the aortic ring, similarly to the 3mensio software. From there, the fluoroscopic angles and ring measurements are made automatically (figure 1). 50 consecutive successful TAVI cases were selected. Measurements had been carried out by Interventional Cardiologists proficient in TAVI using the 3mensio software. Measurements were then carried out in the 3D slicer plugin. The results of the aortic annulus measurements were then compared between the two software methods (perimeter, area, area derived diameter, mean diameter, maximum diameter, minimum diameter), using the t-test for assessing differences. Results A total of 50 measurements carried out with 3mensio software revealed a mean minimum diameter of 21.1 ±2.68 mm; mean maximum diameter of 26.465 ± 2.92 mm, mean perimeter of 75.26 ± 8.62mm and mean area of 442.07 ± 99.63 mm2. Regarding the measurements done with the python plugin, the mean minimum diameter was 21.09 ±2.8mm; mean maximum diameter was 26.47± 3.33 mm; mean perimeter was 75.89 ± 7.86 mm; mean area was 443.98 ±104.45 mm2 and a mean area derived from mean diameter was 236.19± 27.5 mm2. The comparison of the two measurements revealed that only the perimeter was significantly different between the 2 tools (p value <0.001), with an absolute mean difference of 1.21 mm. Conclusions Our plugin produced measurements for aortic annulus dimensions that were highly comparable to those generated by the gold standard 3mensio software, with no significant differences observed for most parameters. The only exception was the perimeter, which showed a statistically significant but small absolute difference of 1.21 mm. Despite being statistically significative, this difference is probably not clinically relevant. The open-source plugin may offer a viable alternative for TAVI sizing, potentially reducing software-related costs and increasing accessibility to TAVI programs.
No takes yet. Share an insight, caveat, or question.
Esteves et al. (2025) conducted an observational in TAVI sizing (n=50). Open-source 3D Slicer python plugin vs. 3mensio software was evaluated on Aortic annulus perimeter (MD 1.21, p=<0.001). An open-source 3D Slicer plugin yielded aortic annulus measurements highly comparable to 3mensio software, with only a small difference in perimeter.