Key result
Reshaping the ventricular patch reduced left ventricular volume from 66 ± 7 mL to a mean of 38 ± 7 mL in porcine hearts.
Why the study?
Does a variably shaped resizable ventricular patch reduce end-diastolic volume and impact cardiac performance in a porcine model of left ventricular aneurysm?
Does a variably shaped resizable ventricular patch reduce end-diastolic volume and impact cardiac performance in a porcine model of left ventricular aneurysm?
A resizable ventricular patch can effectively reduce end-diastolic volume and alter cardiac performance in a preclinical porcine model of left ventricular aneurysm.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for resizable patches in porcine LV aneurysm; leaves open translation to human outcomes.
Objective Endoventricular circular patch plasty is a method used to reconstruct the ventricular cavity in patients with (post) ischemic left ventricular aneurysm or global dilatation. However, late redilatation with mitral regurgitation has been reported, in which postoperative apex shape seems to play an important role. We studied the feasibility of ventricular volume downsizing with a variably shaped patch in porcine hearts. Methods In five in vitro and two acute animal experiments, a dyskinetic aneurysm was simulated with a pericardial insert. Reducing patch surface by changing patch shape diminished end-diastolic volume. In vitro, static end-diastolic volume was determined for each patch shape using volumetry and echocardiography. In the acute animal experiments, preliminary observations of patch behavior in live material were made, and pressure/time relationship, dPdTmax, was registered. Results In vitro, bringing the convex patch into a flat plane reduced LV volume from 66 ± 7 mL (aneurysm) to 49 ± 5 mL. Four of 5 patch shapes further reduced volume to a mean of 38 ± 7 mL (P = 0.03). The in vitro echocardiographic measurements correlated with volumetry findings (r = 0.81). In the acute animal experiments, dPdTmax varied with patch shape, independent of volume changes. Conclusions In this pilot study, in vitro shape configuration of the resizable ventricular patch resulted in a calibrated end-diastolic volume reduction. The data of the two in vivo pilot experiments clearly indicate that change in patch configuration in the situation of more or less unchanged end-diastolic volume had impact on cardiac performance. Future studies must substantiate the results of this observation.
No takes yet. Share an insight, caveat, or question.
Huijgen et al. (2010) studied this question. Reshaping the ventricular patch reduced left ventricular volume from 66 ± 7 mL to a mean of 38 ± 7 mL in porcine hearts.