Key result
Transseptal percutaneous retrieval resolves lattice-tip catheter entrapment in AV valves during VT ablation without long-term harm.
Why the study?
Initial experience with the novel lattice-tip catheter has shown promising results for complex ventricular tachycardias, but safety data regarding catheter-valve interactions remain limited.
First reported cases of lattice-tip catheter entrapment in the atrioventricular valve apparatus during VT ablation, highlighting the mechanical risks of retrograde transaortic access and the utility of transseptal bailout strategies.
Answers are grounded in this paper and cite their sources.
Alerts to entrapment risk with retrograde lattice-tip access in VT ablation; leaves open optimal strategies pending larger experience.
Background The novel lattice-tip catheter (Sphere-9, Medtronic) enables high-resolution mapping and the ability to switch between radiofrequency and pulsed electric field. Initial experience has shown promising results for treating complex ventricular tachycardias (VTs). However, safety data remain limited, particularly regarding catheter–valve interactions. Cases summary We report two cases of mechanical entrapment of the lattice-tip catheter in the atrioventricular valves during VT ablation performed via a retrograde transaortic approach. In both cases, the catheter became entangled within the subvalvular apparatus—specifically, in the tricuspid valve of a patient with congenitally corrected transposition of the great arteries, and in the mitral valve of a patient who had previously undergone mitral repair. Standard retrieval maneuvers were unsuccessful, and resolution required percutaneous bailout strategies, specifically transseptal access with bioptomes or forceps. Both procedures were completed without long-term sequelae. These are the first reported cases of large-footprint catheter entrapment in the atrioventricular valve apparatus, highlighting potential mechanical risks associated with its employment in the ventricular setting. Discussion Our experience emphasizes the importance of preferring transseptal access over retrograde access where feasible, due to its more favorable trajectory and the ability to support safer rescue techniques. Real-time imaging, particularly intracardiac echocardiography, proved essential both for navigation and complication management. While transaortic access remains a viable option, it should be approached with caution, necessitating a thorough understanding of device behavior and a readiness to manage complications. These observations add important safety considerations for the expanded use of novel ablation platforms in patients with complex ventricular anatomy.
No takes yet. Share an insight, caveat, or question.
Ventrella et al. (2026) studied this question. Percutaneous retrieval via transseptal access successfully resolved lattice-tip catheter entrapment in atrioventricular valves after retrograde approach during VT ablation without long-term harm.
Synapse has enriched 10 related papers. Consider them for comparative context: