76.32% of systolic and 69.89% of diastolic BP measurements from home BPMs were within 5 mmHg of reference; over 20% exceeded this difference.
Are home upper-arm oscillometric blood pressure monitors accurate compared to an FDA-certified auscultatory device in outpatient clinical patients?
Over 20% of home blood pressure monitors used by outpatients exhibit clinically significant discrepancies (>5 mmHg) compared to reference standards, highlighting the need for regular device validation.
Absolute Event Rate: 0% vs 0%
This study evaluated the accuracy of a variety of home upper-arm oscillometric blood pressure monitors (BPMs) from 448 participants against the FDA-certified XYZ110 auscultatory device using simultaneous measurements. Descriptive statistics were used to summarize the key findings of the study. Device performance was evaluated referencing international standards (ISO 81060-2:2013), where a mean absolute difference of ≤5 mmHg was defined as "Accurate." In addition, survey questionnaires completed by a subset of participants were analyzed to provide supplementary insights. A total of 448 consecutive outpatient clinical patients attending a routine clinical visit with their BPMs (male 215 and female 233) were eligible to participate in this study. The overall mean age of the participants was 62.35±12.59 years. Most of the BPMs included in this study were of the Omron brand (79.69%, n = 357), followed by Yuwell (5.8%, n = 26) and others. 76.32% of systolic blood pressure (SBP) measurements and 69.89% of diastolic blood pressure (DBP) measurements from BPMs exhibited differences of ≤5 mmHg compared to the reference calibrator. 70.89% (n = 318) were accurate in measuring SBP, and 60.27% (n = 270) were accurate in measuring DBP. Our study revealed that over 20% of BPMs exhibited discrepancies of more than 5 mmHg compared to a reference calibrator. Overall, the Omron U30, HEM-7211, and U10 models demonstrated relatively higher accuracy in blood pressure measurements. Regular patient to patient validation of BPMs is crucial to ensure accurate measurements for daily use.
Zhang et al. (Sun,) reported a other. 76.32% of systolic and 69.89% of diastolic BP measurements from home BPMs were within 5 mmHg of reference; over 20% exceeded this difference.