June 1, 2001Blood Pressure Monitoring

Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor

View Full Paper
Ask AI
Bookmark
Share

Why the study?

Does the new algorithm for the BPM-100 automated oscillometric blood pressure monitor accurately measure blood pressure compared to a mercury sphygmomanometer reference?

Population

85 subjects from an independently conducted validation trial comparing the BPM-100 with a mercury…

Comparison

New algorithm for the BPM-100 automated… vs Reference measurements

Design

Other

Authors

JWJames M WrightGeneral CardiologyGMGurdial S. MattuTPThomas L. PerryUniversity of British Columbia

Discussion

Loading...

Member takes

Overview

BPM-100 accuracy supports clinical consideration; extends device validation literature while leaving outcome data open.

Key Points

  • The study aims to test the accuracy of a new algorithm for the BPM-100 blood pressure monitor using previously collected data.
  • Stored pulse wave and cuff pressure data from 391 measurements were analyzed.
  • A new algorithm was developed from a subset of 136 measurements to enhance accuracy for higher systolic pressures.
  • The larger subset of 255 measurements was utilized to validate the algorithm's accuracy against a mercury sphygmomanometer.
  • Mean difference for systolic BP was -0.16 +/- 5.13 mmHg, with 73.7% within 5 mmHg, 94.9% within 10 mmHg, and 98.8% within 15 mmHg.
  • Mean difference for diastolic BP was -1.41 +/- 4.67 mmHg, with 78.4% within 5 mmHg, 92.5% within 10 mmHg, and 99.2% within 15 mmHg.
  • The algorithm passed the AAMI standard and received 'A' grade in the BHS protocol.

Structured PICO

Does the new algorithm for the BPM-100 automated oscillometric blood pressure monitor accurately measure blood pressure compared to a mercury sphygmomanometer reference?

P
Population
85 subjects (providing 255 measurements) from an independently conducted validation trial comparing the BPM-100(Beta) with a mercury sphygmomanometer
I
Intervention
New algorithm for the BPM-100 automated oscillometric blood pressure monitor
C
Comparator
Reference measurements (mean of two observers using a mercury sphygmomanometer)
O
Outcome
Mean difference +/- SD between the new algorithm BPM-100 and the reference, and percentage of measurements within 5, 10, and 15 mmHgsurrogate

The new algorithm for the BPM-100 automated oscillometric blood pressure monitor meets the AAMI standard and BHS protocol for accuracy.

Cite This Study

Wright et al. (2001) studied this question.

synapsesocial.com/papers/6a209e20e307124fcfcd2f42https://doi.org/10.1097/00126097-200106000-00008

Topics

Hypertension management
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Evaluation of Blood Pressure Measuring Devices1993 · 12 citations
  2. 2Ambulatory Blood Pressure Is Superior to Clinic Blood Pressure in Predicting Treatment-Induced Regression of Left Ventricular Hypertrophy1997 · 596 citations
  3. 3Short report: An outline of the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devices1993 · 522 citations
  4. 4Guidelines for the use of self-blood pressure monitoring2000 · 214 citations
  5. 5The Nellcor N‐101 pulse oximeter1986 · 32 citations