Why the study?
Does the BPM-100(Beta) automated oscillometric blood pressure monitor accurately measure blood pressure compared to standard auscultatory mercury sphygmomanometry?
Population
85 subjects (out of 92 recruited) meeting inclusion criteria
Comparison
BPM-100 automated oscillometric blood pressure… vs Standard auscultatory mercury sphygmomanometry
Design
Cross-sectional, Two trained observers blinded from each other and from the…
Authors
Loading...
Supports BPM-100(Beta) accuracy versus reference; leaves open prospective outcome trials before practice change.
Does the BPM-100(Beta) automated oscillometric blood pressure monitor accurately measure blood pressure compared to standard auscultatory mercury sphygmomanometry?
The BPM-100(Beta) automated oscillometric blood pressure monitor is accurate for office settings, meeting AAMI requirements and achieving a BHS 'A' grade.
Mattu et al. (2001) studied this question.