Key result
Novel zero-fluoroscopy i-Pulse technique achieves acute pulmonary vein isolation across all treated cases.
Case Report (n=3)
No
The i-Pulse technique demonstrates the feasibility of performing zero-fluoroscopy pulsed-field ablation for pulmonary vein isolation using real-time impedance-based contact mapping alongside ICE guidance.
Answers are grounded in this paper and cite their sources.
Supports feasibility of zero-fluoroscopy pulsed-field PVI in three cases; hypothesis-generating and requires prospective validation.
Impedance-guided PulseSelect contact mapping (i-Pulse) adapts EnSite X Contact Index for a circular multielectrode PFA catheter, enabling real-time visualization of electrode-tissue contact without fluoroscopy. With ICE guidance, Δimpedance > 5% identified adequate contact, standardized tagging, and guided applications in three atrial fibrillation cases, achieving acute pulmonary vein isolation without complications.
No takes yet. Share an insight, caveat, or question.
Yamashita et al. (2026) conducted a case report in Atrial fibrillation (n=3). Impedance-guided PulseSelect contact mapping (i-Pulse) was evaluated on Acute pulmonary vein isolation. The i-Pulse technique utilizing EnSite X and ICE guidance successfully achieved complete acute pulmonary vein isolation without complications in all three cases without the use of fluoroscopy.