Key result
Complete pulmonary vein isolation using the Globe multielectrode array catheter was successfully achieved in all 12 veins across 3 patients with atrial fibrillation.
Why the study?
Does the Globe multielectrode array catheter achieve pulmonary vein isolation in patients with atrial fibrillation?
Case Report (n=3)
Does the Globe multielectrode array catheter achieve pulmonary vein isolation in patients with atrial fibrillation?
The novel Globe multielectrode array catheter demonstrated feasibility for achieving complete pulmonary vein isolation in a first-in-human experience of 3 patients with atrial fibrillation.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating in 3 patients; larger trials needed before any clinical role for the Globe catheter.
INTRODUCTION: One question for the technological advancement of catheter ablation of atrial fibrillation (AF) is whether a creative new concept can combine and even improve the diagnostic mapping options of single-tip and basket catheters with the simplicity of the use of balloon catheters for ablation. Herein, we describe the first in-human experience with a single catheter offering such a complete solution. METHODS AND RESULTS: ) with a distal multielectrode array consisting of 16 ribs with 122 gold-plated electrodes was used. Each electrode can ablate, pace, and can measure tissue contact, temperature, current, and intracardiac electrograms. The Globe was deployed and removed without difficulty in all 3 patients. Complete pulmonary vein isolation (PVI) was achieved in all 12 veins. In 10 veins, PVI was achieved with a single placement in front of the respective vein ("single circle isolation"). In one subject, the device was repositioned due to the esophagus location. In the other subject, a single gap was observed after circumferential ablation of the right inferior PV. After precise gap identification, the device was adjusted slightly for improved contact at that region, and reablation resulted in immediate PVI. CONCLUSIONS: PVI isolation could be performed with the new multielectrode array Globe in all 12 PVs offering the option for easy handling and fast "single-shot" PVI. Several continuously updated mapping types from 122 electrodes even in real time during ablation demonstrate the capability to go beyond PVI for voltage mapping plus substrate modification, and for rotor mapping plus rotor ablation.
No takes yet. Share an insight, caveat, or question.
Kottkamp et al. (2017) conducted a case report in Atrial fibrillation (n=3). Globe multielectrode array catheter was evaluated on Complete pulmonary vein isolation (PVI). Complete pulmonary vein isolation using the Globe multielectrode array catheter was successfully achieved in all 12 veins across 3 patients with atrial fibrillation.
Synapse has enriched 10 related papers. Consider them for comparative context: