Key result
CRT-P Safety Mode associated with AV dissociation and worsening heart failure, requiring urgent generator replacement.
Why the study?
Boston Scientific dual-chamber INGENIO family pacemakers or CRT-P devices may initiate safety mode, which makes telemetry unavailable and can lead to adverse clinical consequences.
Safety mode activation in INGENIO family CRT-P devices can cause loss of AV synchronization and worsening heart failure, necessitating urgent generator replacement since mode adjustment becomes impossible.
Answers are grounded in this paper and cite their sources.
Challenges safety of automatic mode switching in INGENIO CRT-P; leaves open need for post-market surveillance and potential design revisions.
Boston Scientific reported that dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-P) may initiate safety mode. Once Safety Mode is activated, telemetry with the programmer becomes unavailable. This case highlights worsening heart failure due to loss of atrioventricular synchronization associated with transition to Safety Mode and VVI pacing. Urgent admission for generator replacement was necessary because mode adjustment was not possible.
No takes yet. Share an insight, caveat, or question.
Osada et al. (2026) studied this question. Transition to Safety Mode in CRT-P can cause AV dissociation and worsen heart failure, requiring urgent generator replacement as mode adjustment is impossible.