Systematic pacing maneuvers and electrocardiogram interpretation rapidly differentiate lead-to-port reversal from mechanical lead failure, preventing inappropriate lead revision.
Systematic pacing maneuvers and ECG analysis can prevent inappropriate lead revision by identifying lead-to-port reversal after CRT-D generator replacement.
Absolute Event Rate: 0% vs 0%
This case highlights how systematic pacing maneuvers and careful electrocardiogram interpretation can rapidly differentiate electrical misconnection from mechanical lead failure, preventing misdiagnosis and inappropriate lead revision.
Şahin et al. (Sun,) reported a other. Systematic pacing maneuvers and electrocardiogram interpretation rapidly differentiate lead-to-port reversal from mechanical lead failure, preventing inappropriate lead revision.