Key result
Stepwise pacing maneuvers identify inadvertent LV-RA port reversal, preventing unnecessary CRT-D lead revision.
Case Report (n=1)
No
Systematic pacing maneuvers and careful ECG interpretation can rapidly differentiate electrical lead-to-port misconnection from mechanical lead failure after CRT-D generator replacement, preventing misdiagnosis and unnecessary lead revision.
Answers are grounded in this paper and cite their sources.
May prevent unnecessary lead revisions after CRT-D replacement; hypothesis-generating for routine pacing maneuvers in device troubleshooting.
BACKGROUND: Lead-to-port misconnections are a rare but clinically significant complication after cardiac device generator replacement and may mimic true lead failure or loss of capture, leading to diagnostic confusion and unnecessary interventions. CASE SUMMARY: A 72-year-old man with ischemic cardiomyopathy and a previously implanted cardiac resynchronization therapy defibrillator presented with worsening dyspnea following elective generator replacement performed at another center. Postprocedural electrocardiogram demonstrated loss of biventricular pacing. Chest radiography confirmed correct lead positions. High-output left ventricle (LV) pacing failed to produce capture, while unipolar LV pacing resulted in surface P waves, and AAI pacing revealed ventricular capture. Stepwise device interrogation excluded lead dislodgement and malfunction, establishing inadvertent LV-right atrium port reversal during generator replacement as the underlying cause. DISCUSSION: This case highlights how systematic pacing maneuvers and careful electrocardiogram interpretation can rapidly differentiate electrical misconnection from mechanical lead failure, preventing misdiagnosis and inappropriate lead revision.
No takes yet. Share an insight, caveat, or question.
Şahin et al. (2026) conducted a case report in Ischemic cardiomyopathy with left ventricle-right atrium lead-to-port reversal after CRT-D generator replacement (n=1). Stepwise device interrogation and pacing maneuvers (unipolar LV pacing and AAI pacing) was evaluated on Diagnosis of inadvertent LV-RA lead-to-port reversal and differentiation from mechanical lead failure. Stepwise pacing maneuvers identified inadvertent left ventricle-right atrium port reversal after CRT-D generator replacement, preventing unnecessary lead revision.
Synapse has enriched 10 related papers. Consider them for comparative context: