Key result
Sternal wire closure traps existing S-ICD lead causing noise and requiring second right sternal lead.
This case highlights the necessity of procedural planning prior to cardiac surgery in patients with existing cardiac implantable electronic devices to prevent complications like jailed leads.
Answers are grounded in this paper and cite their sources.
May warrant pre-surgical planning in S-ICD patients prior to sternotomy; hypothesis-generating for lead management in cardiac surgery.
The subcutaneous-ICD (S-ICD) represents a non-transvenous alternative to traditional transvenous ICDs in patients who do not require pacing and have an elevated risk of infection or are young1. We present the case of an S-ICD sternal lead that was ‘jailed’ after closure of a median sternotomy using sternal wires resulting in noise on the lead. As the lead could not be safely extracted, a second lead was successfully implanted on the right sternal border. This case highlights the necessity of procedural planning prior to cardiac surgery in patients with existing cardiac implantable electronic devices (CIEDs).