Investigating Corneal Trauma in Normal Eyes Using Non-Contact Tonometer, Indicating Minimal Risk.
We investigated the possibility of significant corneal trauma (as revaled by slit lamp observertion of the fluorescein instilled eye), and massage effects following determination of intraocular pressure with the A. O Non Contact tonometer (NCT). Fifteen different, normal human eyes were each aplamated 150 succesive times with the NCT. leading to the conclusion that only minor, superficial corneal epithelial defects sometimes resulted and that, in the line with other studies, the intially higher readins (about I mm), obtained with the NCT, were most likely due to patient apprebension, while the subsequently lower readings represented patient acceptance of the process and were not a result of true aqueous massage. As in an earlier study, we found the instrument variability to be about ± 1 or ± 2 mm and probably due to the subject's own cardiac cycle.